Biomechanical comparison of knotless vs. knotted suture anchors in the acetabular rim with respect to bone density.

Benca Emir, Dlaska Constantin E, DiFranco Matthew D, Hobusch Gerhard M, Windhager Reinhard, Angeli Thomas, von Skrbensky Gobert

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Acetabular labral tears are managed with suture anchors providing good clinical outcomes. Knotless anchors are easier to use and have a quicker insertion time compared to knotted anchors. The purpose of this study was to compare the biomechanical behavior of two different anchor designs (knotted vs. knotless) in ultimate load testing in correlation with bone density in the acetabular rim.

METHODS

Eighteen knotted Bio-FASTak and seventeen knotless PushLock anchors (both: Arthrex Inc., Naples, FL, USA) were inserted in the rims of two human acetabula, with known bone density distribution. The anchors were subjected to load-to-failure tests. Anchors were additionally tested in solid polyurethane foam with defined densities.

FINDINGS: The Bio-FASTak group showed higher survival rates at 1, 2, and 3 mm displacement and was able to withstand significantly higher loads at 3 mm displacement (p = 0.031). There was no statistically significant difference in stiffness (p = 0.087), yield- (p = 0.190), and ultimate load (p = 0.222) between the two groups. Only the PushLock group showed correlation between bone volume over total volume (BV/TV) and stiffness (R = 0.750, p = 0.086) and between BV/TV and yield load (R = 0.838, p = 0.037). Experiments on solid polyurethane foam confirmed the correlation between the mechanical properties and tissue density for the same anchor.

INTERPRETATION: PushLock shows similar biomechanical properties to the Bio-FASTak, but eliminates knot tying and potentially abrasive knots. In addition, biomechanical properties of the PushLock are governed by local bone density.

Для цитирования: Benca Emir, Dlaska Constantin E, DiFranco Matthew D, Hobusch Gerhard M, Windhager Reinhard, Angeli Thomas, von Skrbensky Gobert. Biomechanical comparison of knotless vs. knotted suture anchors in the acetabular rim with respect to bone density.. Clinical biomechanics (Bristol, Avon). 2020. DOI: 10.1016/j.clinbiomech.2020.02.005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.