Biomechanical properties of an intramedullary suture anchor fixation compared to tension band wiring in osteoporotic olecranon fractures- A cadaveric study.

von Keudell Arvind, Kachooei Amir, Mohamadi Amin, Mortensen Sharri J, Okajima Stephen, Egan Jonathan, Weaver MichaelJ, Dyer George S M, Nazarian Ara

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

The aim of the study is to compare three different fixation techniques for transverse olecranon repair in cadaveric osteoporotic bone: (1) current recommended AO tension band technique with K-wire fixation; (2) Suture anchor fixation and (3) Polyester suture fixation.

METHODS

Evaluated with bone densitometry, 7 osteoporotic human elbow specimens were included in the study. A transverse olecranon fracture was reduced anatomically and were fixated first using a K-wire tension band technique, second using two suture anchors, and third using polyester suture. Static simulations of the kinetics associated with active range of motion (AROM) and push up from a chair exercises were performed with cyclic loading using Instron hydraulic testing apparatus. Fracture displacement was measured using videographic analysis. Failure was defined as 2 mm fracture displacement.

RESULTS

The biomechanical analysis found no statistical difference in displacements between the three fixation methods when testing AROM. In simulated push-up exercises, polyester suture fixation failed after 17 cycles and had significantly higher displacement compared to the other two methods. No difference between the K-Wire fixation versus Suture anchor fixation was observed, p = 0.162.

CONCLUSION

Suture anchor fixation might be a viable surgical treatment option for osteoporotic transverse elbow fractures in geriatric patients.

Для цитирования: von Keudell Arvind, Kachooei Amir, Mohamadi Amin, Mortensen Sharri J, Okajima Stephen, Egan Jonathan, Weaver MichaelJ, Dyer George S M, Nazarian Ara. Biomechanical properties of an intramedullary suture anchor fixation compared to tension band wiring in osteoporotic olecranon fractures- A cadaveric study.. Journal of orthopaedics. 2019. DOI: 10.1016/j.jor.2019.08.002.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.