OBJECTIVE

To compare 2 intracorporeal suturing techniques among surgical residents.

METHODS

Surgical residents at an academic teaching hospital were invited to participate. All participants completed 2 intracorporeal suturing trials in a laparoscopic trainer using either endoscopic needle drivers (END) or an automated endoscopic suturing device (ESD), with 14 days between trials. A standardized, descriptive video of each intracorporeal suturing technique was provided 7 days before the respective trial. Time to perform suturing was recorded, and participants completed a National Aeronautics and Space Administration Task Load Index (NASA-TLX) questionnaire following each trial. Suture lines were subjectively blindly scored using a standardized questionnaire by 3 surgeons.

RESULTS

10 surgical residents across 4 years of training were enrolled. Mean time to complete the suturing trial was 5.72 ± 1.49 minutes for END and 6.00 ±2.14 minutes for ESD (P = .76). The mean NASA-TLX score was 44.60 ± 19.02 for END and 44.33 ± 22.22 for ESD (P = .83). Median suture evaluation score was 11 (IQR, 8 to 11) for END and 11 (IQR, 9 to 11) for ESD (P = .13). Mean time to perform suturing was significantly different across years of training in residency for both ESD and END (P = .006).

CONCLUSIONS

In this ex vivo model, an automated ESD provided no benefit for ease of performing intracorporeal suturing among surgical residents.

CLINICAL RELEVANCE: Surgical residents can perform intracorporeal suturing with knotless barbed suture using either ESD or END. Further evaluation of the learning curve of each instrument for novice surgeons is warranted.

Для цитирования: Solari Francesca P, Williams Rachel W, Case J Brad, Ham Kathleen, Colee James C. Use of an automated endoscopic suturing device does not improve mental workload or time of intracorporeal suturing among surgical residents.. American journal of veterinary research. 2026. DOI: 10.2460/ajvr.26.04.0173.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.