MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.radonc.2023.109577

Медицинская библиотека · Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology · 2023

Characterizing the interplay of treatment parameters and complexity and their impact on performance on an IROC IMRT phantom using machine learning.

Mehrens Hunter, Molineu Andrea, Hernandez Nadia, Court Laurence, Howell Rebecca, Jaffray David, Peterson Christine B, Pollard-Larkin Julianne, Kry Stephen F

СтатьяПоддержка NIH: внешняя программаПеревод пока не подготовлен

Аннотация

AIM OF THE STUDY: To elucidate the important factors and their interplay that drive performance on IMRT phantoms from the Imaging and Radiation Oncology Core (IROC).

METHODS

IROC's IMRT head and neck phantom contains two targets and an organ at risk. Point and 2D dose are measured by TLDs and film, respectively. 1,542 irradiations between 2012-2020 were retrospectively analyzed based on output parameters, complexity metrics, and treatment parameters. Univariate analysis compared parameters based on pass/fail, and random forest modeling was used to predict output parameters and determine the underlying importance of the variables.

RESULTS

The average phantom pass rate was 92% and has not significantly improved over time. The step-and-shoot irradiation technique had significantly lower pass rates that significantly affected other treatment parameters' pass rates. The complexity of plans has significantly increased with time, and all aperture-based complexity metrics (except MCS) were associated with the probability of failure. Random forest-based prediction of failure had an accuracy of 98% on held-out test data not used in model training. While complexity metrics were the most important contributors, the specific metric depended on the set of treatment parameters used during the irradiation.

CONCLUSION

With the prevalence of errors in radiotherapy, understanding which parameters affect treatment delivery is vital to improve patient treatment. Complexity metrics were strongly predictive of irradiation failure; however, they are dependent on the specific treatment parameters. In addition, the use of one complexity metric is insufficient to monitor all aspects of the treatment plan.

Для цитирования: Mehrens Hunter, Molineu Andrea, Hernandez Nadia, Court Laurence, Howell Rebecca, Jaffray David, Peterson Christine B, Pollard-Larkin Julianne, Kry Stephen F. Characterizing the interplay of treatment parameters and complexity and their impact on performance on an IROC IMRT phantom using machine learning.. Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology. 2023. DOI: 10.1016/j.radonc.2023.109577.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.