DOI 10.3760/cma.j.issn.0376-2491.2019.45.005

Медицинская библиотека · Zhonghua yi xue za zhi · 2019

[Application of real-time image fusion technique in transjugular intrahepatic portosystemic shunt].

Zhang W W, Ren J Z, Zhou P L, Ding P X, Chen P F, Li F Z, Kuang D L, Han X W

СтатьяПеревод пока не подготовлен

Аннотация

Objective

To assess the feasibility and value of real-time image fusion technique guiding the procedure of transjugular intrahepatic portosystemic shunt(TIPS).

Methods

From July 2017 to May 2018,a total of 48 consecutive patients complicated by portal venous hypertension due to cirrhosis who underwent TIPS were prospectively allocated into two groups that 27 cases underwent normal TIPS and 21 cases underwent image fusion guided TIPS. There were 25 males and 23 females with a mean age of 29-74(51±10) years. The differences of portal vein(PV) between image fusion angiographyand digital subtraction angiography(DSA), and the times of puncture PV, X-ray exposure dose and exposure time and contrast agent amount of all cases were collected and analyzed.

Results

The longitudinal and traverse difference of PV between image fusion angiography and DSA were 1.7-2.5(2.1±0.2) mm and 0.9-1.8(1.4±0.3) mm, respectively.The times of puncture PV, X-ray exposure time and dose, and contrast agent amount between normal TIPS group and image fusion guided TIPS group were 1-7(3.8±0.6) times vs 1-3(2.0±0.6) times, 41-63(53±8)min vs 27-42(35±5) min, 513-787(644±96) mGy vs 357-524(423±59) mGy,102-196(151±23) ml vs 87-145(105±14) ml(all P<0.05), respectively.

Conclusions

There are minor differences between image fusion angiography of PV and DSA. Real-time image fusion guided TIPS is feasible and valuable to reduce intraprocedural X-ray exposure time and dose and contrast agent amount of TIPS.

Для цитирования: Zhang W W, Ren J Z, Zhou P L, Ding P X, Chen P F, Li F Z, Kuang D L, Han X W. [Application of real-time image fusion technique in transjugular intrahepatic portosystemic shunt].. Zhonghua yi xue za zhi. 2019. DOI: 10.3760/cma.j.issn.0376-2491.2019.45.005.