Development of a clinical workflow tool to enhance the detection of vertebral fractures: accuracy and precision evaluation.

Brett Alan, Miller Colin G, Hayes Curtis W, Krasnow Joel, Ozanian Takouhi, Abrams Ken, Block Jon E, van Kuijk Cornelis

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

STUDY

DESIGN

Image analysis model development.

OBJECTIVE

The objective of this study was to develop a novel clinical workflow tool that uses model-based shape recognition technology to allow efficient, semiautomated detailed annotation of each vertebra between T4 and L4 on plain lateral radiographs.

SUMMARY OF BACKGROUND DATA: Identification of prevalent vertebral fractures, especially when not symptomatic, has been problematic despite their importance. There is a recognized need to increase the opportunities to detect vertebral fractures so that clinically beneficial therapeutic interventions can be initiated.

METHODS

Radiographs obtained from 165 subjects in the Canadian Multicenter Osteoporosis Study (CaMos) were used to construct a vertebral shape model of the vertebral column from T4 to L4 using a statistical learning technique, as well as to estimate the accuracy and precision of this automated software tool for vertebral shape analysis. Radiographs showing scoliosis greater than 15 degrees were excluded.

RESULTS

Vertebral contours defined by 95 points per vertebra, represented by 79,895 points in total, were assessed on 841 individual vertebrae. The mean absolute accuracy error calculated over each vertebra in each test image was 1.06 +/- 1.2 mm. This value corresponded to an average 3.4% of vertebral height. The mean precision error, reflecting interobserver variability, per vertebra of the resulting annotations was 0.61 +/- 0.73 mm. This value corresponded to an average 2.3% of vertebral height. Accuracy and precision error estimates did not differ notably by vertebral level.

CONCLUSION

The results of the current study indicate that statistical modeling can provide a robust tool for the accurate and precise semiautomated annotation of vertebral body shape from T4 to L4 in patients who do not have scoliosis greater than 15 degrees . This method may prove useful as a clinical workflow tool to aid the physician in vertebral fracture assessment and might contribute to decision-making about pharmacologic treatment of osteoporosis.

Для цитирования: Brett Alan, Miller Colin G, Hayes Curtis W, Krasnow Joel, Ozanian Takouhi, Abrams Ken, Block Jon E, van Kuijk Cornelis. Development of a clinical workflow tool to enhance the detection of vertebral fractures: accuracy and precision evaluation.. Spine. 2009. DOI: 10.1097/brs.0b013e3181b2eb69.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.