DOI 10.1016/0735-1097(93)90439-8
Implantation and recovery of temporary metallic stents in canine coronary arteries.
Аннотация
OBJECTIVES
The purpose of this study was to test the feasibility of implanting and retrieving a heat-activated recoverable temporary stent and to determine its effect on the angiographic, gross and histologic appearance of a normal coronary artery wall.
BACKGROUND
Permanent coronary stenting is associated with a significant incidence of thrombosis, bleeding and vascular complications. These may be avoided by temporarily stenting for a period of hours to several days.
METHODS
Seventy-eight stents constructed from the shape-memory nickel-titanium alloy nitinol were deployed by balloon expansion in the coronary arteries of 28 dogs and left in place for up to 6 months. Thirty minutes to 1 week after implantation, 70 stents were recovered by flushing the coronary arteries with 3 to 5 ml of 75 degrees C lactated Ringer solution, with collapse of the stent over a recovery catheter and subsequent withdrawal.
RESULTS
All stents were successfully recovered and removed percutaneously. Mean vessel diameter after stenting was 12 +/- 6% (p < 0.05) greater than baseline diameter. Mean vessel diameter after stent removal remained enlarged (6 +/- 3%, p < 0.05). No angiographic or gross evidence of thrombosis, dissection, embolization, migration or spasm was associated with implantation or recovery. Microscopic examination revealed minor intimal injury in 40 segments (51%). Microscopic focal medial necrosis was associated with mural platelet-fibrin thrombus in 23 stented segments (29%) and media was interrupted in 7 (9%).
CONCLUSIONS
This study demonstrates the feasibility of a new method of temporary stenting that uses the thermoelastic properties of nitinol to permit reliable recovery of the stent in normal canine coronary arteries.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.