Impact of electrode-tissue proximity on formation of pulsed field ablation lesions: Using multi-electrode impedance measurements.

Groen Marijn H A, Fish Jeffrey M, Loh Peter, van Es René

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

For pulsed field ablation (PFA), the effect of electrode-tissue contact or proximity remains unclear.

OBJECTIVE

In this porcine study, the effect of electrode-tissue proximity was studied after 3 weeks of follow-up, using a custom device in which the electrode-tissue distance could be varied for every electrode.

METHODS

An impedance-based method was used to determine electrode-tissue proximity or contact. During a bench experiment, the relation between contact value per electrode and electrode-tissue distance was studied. In addition, in 10 pigs epicardial ablation using a custom 6-polar linear ablation device was performed using bipolar, biphasic deliveries on randomized locations on the heart. After 3 weeks, macroscopic analysis was used to compare lesion width for left vs right ventricle. Based on histology, the lesion area and depth were related to electrode-tissue contact.

RESULTS

Analysis of 96 bench measurements showed a decrease in tissue-proximity index with increasing distance. Lesions on the right ventricle were wider compared to the lesions on the left ventricle, 9.0 ± 1.9 mm vs 6.5 ± 1.0 mm (P < .001), respectively. A total of 162 lesions showed that a higher tissue-proximity index was associated with deeper and larger lesions; lesion depth was 1.3 ± 0.5 mm vs 1.5 ± 0.5 mm (P = .016) and lesion area 6.4 ± 2.7 mm2 vs 8.3 ± 2.9 mm2 (P < .001) for low vs high indices respectively.

CONCLUSION

This study shows that higher electrode-tissue proximity may result in deeper PFA lesions, and is, therefore, an important parameter in improving PFA lesions. Therefore, confirmation of tissue contact using impedance-based measurements may improve PFA ablation efficacy in a clinical setting.

Для цитирования: Groen Marijn H A, Fish Jeffrey M, Loh Peter, van Es René. Impact of electrode-tissue proximity on formation of pulsed field ablation lesions: Using multi-electrode impedance measurements.. Heart rhythm O2. 2025. DOI: 10.1016/j.hroo.2025.05.026.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.