Pulsed electric field for pulmonary vein isolation in paroxysmal atrial fibrillation: A multicenter experience comparing pentaspline and lattice-tip catheters.

Almorad Alexandre, Scacciavillani Roberto, Vetta Giampaolo, Lyne Jonathan, Rodriguez Jesus, Zaher Wael, Doundoulakis Ioannis, Sorgente Antonio, Marcon Lorenzo, Pannone Luigi, Crinion Derek, Tuohy Stephen, Nakasone Kazutaka, Pasara Vedran, Del Monte Alvise, Stroker Erwin, Bala Gezim, Audiat Charles, Sieira Juan, Combes Stephane, Letsas Konstantinos P, Kariki Ourania, Efremidis Michael, Boveda Serge, Sarkozy Andrea, Della Rocca Domenico Giovanni, de Asmundis Carlo, Chierchia Gian-Battista

СтатьяПеревод в очереди

Аннотация

BACKGROUND

We conducted a multicenter study to compare procedural and 1-year outcomes of pulmonary vein isolation (PVI) using lattice-tip (LT) vs pentaspline (PS) pulsed field ablation catheters in patients with paroxysmal atrial fibrillation.

OBJECTIVE

This study aimed to compare procedural metrics and arrhythmia-free survival during the 1-year follow-up.

METHODS

Procedures were performed between August 2023 and December 2024. A propensity score matching technique was adopted (LT-to-PS ratio 1:2).

RESULTS

Among 447 patients with paroxysmal atrial fibrillation undergoing first-time PVI with either an LT (n = 74) or a PS catheter (n = 391), propensity score matching yielded 50 (LT group) and 100 patients (PS group), respectively. No differences were reported for first-pass isolation per patient (96% [LT group] vs 97% [PS group]; P = .75) and per pulmonary vein (98.9% [LT group] vs 99.2% [PS group]; P = .75). The LT group showed longer procedural (56 vs 50; P < .0001) and dwelling times (40 vs 30; P < .0001) but reduced fluoroscopy times (5 vs 10; P < .0001). No differences were reported for overall (2% vs 2%; P = 1.0) and major complications (0% vs 0%; P = 1.0). At 1-year follow-up, similar freedom from atrial tachyarrhythmia (86.2% [LT group] vs 83% [PS group]; P = .59) and atrial fibrillation (90.7% [LT group] vs 85.8% [PS group]; P = .41) was observed.

CONCLUSION

First-time PVI using an LT or a PS pulsed field ablation catheter showed high efficacy and 1-year freedom from atrial tachyarrhythmias, without differences between groups. The LT catheter yielded longer procedural and dwelling times, but a significant reduction in fluoroscopy time.

Для цитирования: Almorad Alexandre, Scacciavillani Roberto, Vetta Giampaolo, Lyne Jonathan, Rodriguez Jesus, Zaher Wael, Doundoulakis Ioannis, Sorgente Antonio, Marcon Lorenzo, Pannone Luigi, Crinion Derek, Tuohy Stephen, Nakasone Kazutaka, Pasara Vedran, Del Monte Alvise, Stroker Erwin, Bala Gezim, Audiat Charles, Sieira Juan, Combes Stephane, Letsas Konstantinos P, Kariki Ourania, Efremidis Michael, Boveda Serge, Sarkozy Andrea, Della Rocca Domenico Giovanni, de Asmundis Carlo, Chierchia Gian-Battista. Pulsed electric field for pulmonary vein isolation in paroxysmal atrial fibrillation: A multicenter experience comparing pentaspline and lattice-tip catheters.. Heart rhythm O2. 2026. DOI: 10.1016/j.hroo.2026.04.014.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.