BACKGROUND

Mitral valve repair using expanded polytetrafluoroethylene sutures to replace mitral chordae tendineae is a well-established procedure. However, the incidence of neo-chordae failure causing recurrent mitral regurgitation is not well defined.

METHODS

We have reviewed the reported cases of complications after mitral valve repair related to the use of neo-chordae. This study was mainly carried out through PubMed, Medline, and Google Chrome websites.

RESULTS

We have identified a total of 26 patients presenting with rupture of polytetrafluoroethylene neo-chordae, mostly being described as isolated cases. Few other cases of recurrent mitral regurgitation with hemolysis were found, where reoperation was not caused by neo-chordal failure but most likely by technical errors. At pathological investigation the findings were substantially similar in all reported cases. The neo-chordae retained their length and pliability, became covered with host tissue and rupture was mainly related to suture size. Mild calcification was observed not interfering with chordal function; chordal infection did never occur.

CONCLUSIONS

The use of artificial neo-chordae provides excellent late results with durable mitral valve repair stability. Chordal rupture may occur late postoperatively leading to reoperation because of recurrent mitral regurgitation. Despite its rarity, this potential complication should not be overlooked during follow-up of patients after mitral valve repair using artificial neo-chordae.

Для цитирования: Vendramin Igor, Milano Aldo D, Pucci Angela, Lechiancole Andrea, Sponga Sandro, Bortolotti Uberto, Livi Ugolino. Artificial chordae for mitral valve repair.. Journal of cardiac surgery. 2022. DOI: 10.1111/jocs.16937.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.