Pulsed Field Ablation for Atrial Fibrillation in Australia: A Large Tertiary Centre Analysis.

Ha Francis J, Prinsloo Duron, Xu Edwin, Han Hui Chen, Adam David, Healy Stewart, Alasti Mohammad, Machado Colin, Bittinger Logan, Cheng Sing Huey, Adhikari Jagat, Nerlekar Nitesh, Brown Adam J, Alison Jeffrey, Kotschet Emily

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Pulsed field ablation (PFA) is increasingly used for catheter ablation of atrial fibrillation (AF). Limited real-world data exists regarding safety and efficacy of PFA in Australia.

METHOD: We conducted a retrospective analysis of consecutive patients undergoing de novo PFA for paroxysmal or persistent AF at a large tertiary centre using a pentaspline catheter (FARAPULSE, Boston Scientific) from 2022 to 2025. Baseline demographics, procedural characteristics, procedural complications, clinical outcomes and repeat procedures were collected. Arrhythmia recurrence was defined as >30 seconds of AF, atrial flutter or atrial tachycardia at follow-up with no blanking period.

RESULTS

Overall, 564 consecutive patients (median age 64 years, 35% female) underwent de novo PFA for AF during the study period. Most patients had paroxysmal AF (63%). Pulmonary vein (PV) isolation was performed in 563 of 564 patients; posterior wall (PW) isolation was performed in 41% and superior vena cava ablation in 3%. Median total applications were 58 (interquartile range [IQR] 50-70). Median total procedural time was 63 minutes (IQR 51-76). Major procedural complications occurred in 1.8% (10/564) patients and minor procedural complications occurred in 12.6% (71/564), mostly because of minor vascular access complications (8.2%). Six (6)-month freedom from arrhythmia was 72%; cardioversions were performed in 5.3% and 2.7% had arrhythmia-related hospitalisation. In 35 patients (6.2%) who underwent repeat procedure, PV and PW durability was 45% and 32%, respectively.

CONCLUSIONS

PFA for AF is reasonably safe with acceptable clinical outcomes. These real-world data provide clinicians with potential expected clinical outcomes for patients referred for PFA in Australia.

Для цитирования: Ha Francis J, Prinsloo Duron, Xu Edwin, Han Hui Chen, Adam David, Healy Stewart, Alasti Mohammad, Machado Colin, Bittinger Logan, Cheng Sing Huey, Adhikari Jagat, Nerlekar Nitesh, Brown Adam J, Alison Jeffrey, Kotschet Emily. Pulsed Field Ablation for Atrial Fibrillation in Australia: A Large Tertiary Centre Analysis.. Heart, lung & circulation. 2026. DOI: 10.1016/j.hlc.2026.04.125.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.