Evaluation of the impact of extracorporeal membrane oxygenation (ECMO) oxygenator shunt flow on neonatal hemolysis: An in vitro study.

Nemeh Christopher, Varner Caleb, Schmoke Nicholas, Trager Elon, Wu Yeu Sanz, Brewer Michael, Cain-Trivette Caitlin, Aw Nikki, Francis Richard O, Vargas Chaves Diana, Cheung Eva W, Middlesworth William

СтатьяПеревод пока не подготовлен

Аннотация

IntroductionExtracorporeal membrane oxygenation (ECMO) is utilized in critically ill neonates with severe cardiopulmonary failure. Hemolysis is a potential complication and is associated with significantly increased morbidity and mortality. The etiology of hemolysis in neonates is multifactorial, including shear forces generated by the ECMO pump, higher flow resistance from smaller tubing and smaller cannulas, the oxygenator, and other patient factors. Centrifugal pumps and oxygenators commonly have shunts with partially occluding clamps to regulate blood flow. We hypothesized that these clamps are significant contributors to hemolysis.MethodAn in vitro study was conducted with three identical ECMO circuits containing an integrated polymethylpentene (PMP) oxygenator and centrifugal pump (Cardiohelp HLS 5.0) and 1/4″ arteriovenous (AV) loop tubing. The circuits were primed with equal components, including expired ABO-compatible packed red blood cells (pRBCs), 25% albumin, 5% albumin, sodium bicarbonate, heparin, and calcium chloride. Circuit A had a completely occluded shunt. Circuit B had a partially occluded shunt, allowing 500 mL/min of shunt flow back to the oxygenator. Circuit C had a fully open shunt, generating 1000 mL/min of shunt flow back to the oxygenator. Plasma-free hemoglobin values were measured serially over 5 days.ResultsBaseline plasma-free hemoglobin levels were equal in all three circuits. Circuit C had the greatest increase in plasma-free hemoglobin daily (26.3 mg/dL/day) compared to Circuit A and Circuit B, which were 14.1 mg/dL/day and 12.9 mg/dL/day, respectively.ConclusionsOur data suggests that partially occluding clamps are not a significant contributor to hemolysis; rather, increased flow through the oxygenator significantly increased the rate of hemolysis in neonatal ECMO circuits.

Для цитирования: Nemeh Christopher, Varner Caleb, Schmoke Nicholas, Trager Elon, Wu Yeu Sanz, Brewer Michael, Cain-Trivette Caitlin, Aw Nikki, Francis Richard O, Vargas Chaves Diana, Cheung Eva W, Middlesworth William. Evaluation of the impact of extracorporeal membrane oxygenation (ECMO) oxygenator shunt flow on neonatal hemolysis: An in vitro study.. Perfusion. 2025. DOI: 10.1177/02676591251322000.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.