Use of research-grade radioisotopes for in-human recovery and survival studies.
Аннотация
BACKGROUND AND
OBJECTIVES
Red blood cells (RBCs) and platelets are evaluated for in vivo quality using recovery and survival measurements in healthy volunteers. Radiolabeling is the standard method for tracking blood cells post-transfusion. The supply of chromium-51 (51Cr) manufactured under Good Manufacturing Practices (GMP) has been challenging, and in 2024, the only remaining supplier ceased production. Now, only research-grade 51Cr is available, requiring risk control measures. Simultaneously, the Pyrophosphate (PYP) kits used to condition RBCs for technetium-99m (99mTc) labeling are also no longer available. This report summarizes validations performed leading to reinfusions of blood cells labeled with a new kit for 99mTc RBC labeling and research-grade 51Cr.
MATERIALS AND
METHODS
Three validations were conducted to confirm: (1) labeling efficiency of an alternative kit for 99mTc labeling RBCs, (2) safety and labeling efficiency using research-grade 51Cr, and (3) the ability of the Pyrosate®-Rapid Endotoxin Detection (PRED) kit to detect pyrogens in research-grade material and in end-products for in vivo use.
RESULTS
Labeling efficiency of RBCs with 99mTc using the Ultratag™ RBC kit was >98% and demonstrated consistency in labeling across different sample preparation conditions. Filtration of the radioisotope stock and endotoxin testing of the infusate served as effective risk mitigation steps, supporting safe use of research-grade 51Cr, with labeling efficiency >98%.
CONCLUSIONS
We validated the use of research-grade 51Cr, an updated kit for 99mTc labeling, and endotoxin testing for in vivo reinfusion of research-grade 51Cr labeled blood cells, and have infused >30 research participants with these changes under US FDA-approved protocols.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.