DOI 10.1016/j.jtcvs.2003.09.055

Медицинская библиотека · The Journal of thoracic and cardiovascular surgery · 2004

The Impella Recover microaxial left ventricular assist device reduces mortality for postcardiotomy failure: a three-center experience.

Siegenthaler M P, Brehm K, Strecker T, Hanke T, Nötzold A, Olschewski M, Weyand M, Sievers H, Beyersdorf F

Оценочное исследованиеСтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

We evaluated patient outcomes and complications associated with the microaxial Impella Recover left ventricular assist device (Impella Cardiosystems AG, Aachen, Germany) for postcardiotomy low-output syndrome. This low-cost device is inserted across the aortic valve through a 10-mm vascular graft sewn to the ascending aorta.

METHODS

Impella patients were compared with 198 patients treated with an intraoperative intra-aortic balloon pump between January 2000 and December 2002. Three risk scores were used: the Hausmann score, the Texas Heart Institute score, and the Cleveland intensive care unit score. Between September 2001 and March 2003, 24 patients were treated with the Impella Recover for low-output syndrome. Before device insertion, 21 could not be separated from cardiopulmonary bypass, and 3 had postoperative hemodynamic instability despite high-dose catecholamines. Sixteen were treated with the Impella and intra-aortic balloon pump and 8 with the Impella alone (no intra-aortic balloon pump because of peripheral vascular disease or because deemed unnecessary).

RESULTS

No technical problems with device insertion occurred. Pump flow was 3.3 +/- 0.7 L/min at 28,000 +/- 4500 RPM. Support time was 61 +/- 56 hours (range, 7-228 hours). Four devices required repositioning. One device failed (leaking purge line) and was removed. Hemolysis was minimal (lactate dehydrogenase levels of 540 +/- 260 U/dL for Impella survivors). Mortality for Impella patients was 54% (13/24), similar to that for high-risk intra-aortic balloon pump patients (Hausmann score > or =2 [57%], intensive care unit score > or =2 [51%], Texas Heart Institute score > or =0.75 [55%], and cardiac index < or =2.3 [45%]). Cardiac output data were available in 19 Impella patients. Impella patients able to increase their cardiac output to 1 L/min or more above the pump flow of the Impella Recover had a 10% (1/10) mortality, versus 88% (8/9) in patients with a residual cardiac function of 1 L/min or less (P =.001). Comparison of high-risk intra-aortic balloon pump patients with Impella patients with residual cardiac function of 1 L/min or more showed a significant reduction in mortality, regardless of the high-risk definition used. Residual cardiac function was the strongest predictor of survival in Impella patients.

CONCLUSIONS

The Impella Recover device provides 3 to 4 L/min flow. It improves survival in patients with low-output syndrome if the heart is able to pump 1 L/min or more above device flow.

Для цитирования: Siegenthaler M P, Brehm K, Strecker T, Hanke T, Nötzold A, Olschewski M, Weyand M, Sievers H, Beyersdorf F. The Impella Recover microaxial left ventricular assist device reduces mortality for postcardiotomy failure: a three-center experience.. The Journal of thoracic and cardiovascular surgery. 2004. DOI: 10.1016/j.jtcvs.2003.09.055.

Предметные рубрики MeSH