Thromboembolization associated with sudden increases in flow in a coronary stent ex vivo shunt model.

Sukavaneshvar S, Zheng Y, Rosa G M, Mohammad S F, Solen K A

СтатьяПеревод пока не подготовлен

Аннотация

To observe the dynamics of thromboembolism (TE) in an animal model, a carotid-carotid arterial ex vivo shunt was developed. A coronary stent deployed in a 3.5 mm polyvinylchloride (PVC) tubing segment was used as a model device in the shunt. A light-scattering microemboli detector monitored the embolic content of the blood flowing through the shunt at 50-150 ml/min as determined by a clamp-on ultrasound flow probe. The stent was found to actively develop thrombi and release emboli for 1-3 hours when the activated clotting time (ACT) was maintained between 125 and 150 seconds. The shunt flow rate fluctuated considerably (from 50 to 150 ml/min) depending on the animal's activity. When the time profile of this fluctuating flow rate was super imposed on the time profile of embolization, it was noted that sudden increases in flow rate were associated with incidents of embolization. Statistically, sudden flow rate increases of 100% or more were accompanied by embolic events 95% of the time (p < 0.01). Based on the results of this study, it was postulated that the increased embolization may be due to the fluid forces associated with accelerating flow. To explore this postulate, in vitro studies were conducted to compare the effects of pulsatile flow with steady flow on stent induced TE. Results of this study suggested a significant increase (100%) in both stent thrombosis and embolism during pulsatile flow compared with steady flow.

Для цитирования: Sukavaneshvar S, Zheng Y, Rosa G M, Mohammad S F, Solen K A. Thromboembolization associated with sudden increases in flow in a coronary stent ex vivo shunt model.. ASAIO journal (American Society for Artificial Internal Organs : 1992). 2000. DOI: 10.1097/00002480-200005000-00012.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.