Медицинская библиотека · Brachytherapy · 2018

Can MRI-only replace MRI-CT planning with a titanium tandem and ovoid applicator?

Harkenrider Matthew M, Patel Rakesh, Surucu Murat, Chinsky Bonnie, Mysz Michael L, Wood Abbie, Ryan Kelly, Shea Steven M, Small William, Roeske John C

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE/OBJECTIVE(S): To evaluate dosimetric differences between MRI-only and MRI-CT planning with a titanium tandem and ovoid applicator to determine if all imaging and planning goals can be achieved with MRI only.

MATERIALS/

METHODS

We evaluated 10 patients who underwent MRI-CT-based cervical brachytherapy with a titanium tandem and ovoid applicator. High-risk clinical target volume and organs at risk were contoured on the 3D T2 MRI, which were transferred to the co-registered CT, where the applicator was identified. Retrospectively, three planners independently delineated the applicator on the axial 3D T2 MRI while blinded to the CT. Identical dwell position times in the delivered plan were loaded. Dose-volume histogram parameters were compared to the previously delivered MRI-CT plan.

RESULTS

There were no significant differences in dose to D90 or D98 of the high-risk clinical target volume with MRI vs. MRI-CT planning. MRI vs. MRI-CT planning resulted in mean D0.1cc bladder of 8.8 ± 3.4 Gy vs. 8.5 ± 3.2 Gy (p = 0.29) and D2cc bladder of 6.2 ± 1.4 Gy vs. 6.0 ± 1.4 Gy (p = 0.33), respectively. Mean D0.1cc rectum was 5.7 ± 1.2 Gy vs. 5.3 ± 1.2 Gy (p = 0.03) and D2cc rectum 4.0 ± 0.8 Gy vs. 4.2 ± 1.0 Gy (p = 0.18), respectively. Mean D0.1cc sigmoid was 5.2 ± 1.3 Gy vs. 5.4 ± 1.6 Gy (p = 0.23) and D2cc sigmoid 3.9 ± 1.0 Gy vs. 4.0 ± 1.1 Gy (p = 0.18), respectively.

CONCLUSION

There were no clinically significant dosimetric differences between the MRI and MRI-CT plans. This study demonstrates that cervical brachytherapy with a titanium applicator can be planned with MRI alone, which is now our clinical standard.

Для цитирования: Harkenrider Matthew M, Patel Rakesh, Surucu Murat, Chinsky Bonnie, Mysz Michael L, Wood Abbie, Ryan Kelly, Shea Steven M, Small William, Roeske John C. Can MRI-only replace MRI-CT planning with a titanium tandem and ovoid applicator?. Brachytherapy. 2018. DOI: 10.1016/j.brachy.2018.05.010.