Rogers Toby, Greenspun Benjamin C, Weissman Gaby, Torguson Rebecca, Craig Paige, Shults Christian, Gordon Paul, Ehsan Afshin, Wilson Sean R, Goncalves John, Levitt Robert, Hahn Chiwon, Parikh Puja, Bilfinger Thomas, Butzel David, Buchanan Scott, Hanna Nicholas, Garrett Robert, Buchbinder Maurice, Asch Federico, Garcia-Garcia Hector M, Okubagzi Petros, Ben-Dor Itsik, Satler Lowell F, Waksman Ron

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVES

The aim of this study was to evaluate the feasibility of coronary access and aortic valve reintervention in low-risk patients undergoing transcatheter aortic valve replacement (TAVR) with a balloon-expandable transcatheter heart valve (THV).

BACKGROUND

Younger, low-risk TAVR patients are more likely than older, higher risk patients to require coronary angiography, percutaneous coronary intervention, or aortic valve reintervention, but their THVs may impede coronary access and cause coronary obstruction during TAVR-in-TAVR.

METHODS

The LRT (Low Risk TAVR) trial (NCT02628899) enrolled 200 subjects with symptomatic severe aortic stenosis to undergo TAVR using commercially available THVs. Subjects who received balloon-expandable THVs and who had 30-day cardiac computed tomographic scans were included in this study. In a subgroup, the feasibility of intentional THV crimping on the delivery catheter to pre-determine commissural alignment was tested.

RESULTS

In the LRT trial, 168 subjects received balloon-expandable THVs and had 30-day cardiac computed tomographic scans, of which 137 were of adequate image quality for analysis. The most challenging anatomy for coronary access (THV frame above and commissural suture post in front of a coronary ostium) was observed in 9% to 13% of subjects. Intentional THV crimping did not appear to meaningfully affect commissural alignment. The THV frame extended above the sinotubular junction in 21% of subjects, and in 13%, the distance between the THV and the sinotubular junction was <2 mm, signifying that TAVR-in-TAVR may not be feasible without causing coronary obstruction.

CONCLUSIONS

TAVR may present challenges to future coronary access and aortic valve reintervention in a substantial number of low-risk patients.

Для цитирования: Rogers Toby, Greenspun Benjamin C, Weissman Gaby, Torguson Rebecca, Craig Paige, Shults Christian, Gordon Paul, Ehsan Afshin, Wilson Sean R, Goncalves John, Levitt Robert, Hahn Chiwon, Parikh Puja, Bilfinger Thomas, Butzel David, Buchanan Scott, Hanna Nicholas, Garrett Robert, Buchbinder Maurice, Asch Federico, Garcia-Garcia Hector M, Okubagzi Petros, Ben-Dor Itsik, Satler Lowell F, Waksman Ron. Feasibility of Coronary Access and Aortic Valve Reintervention in Low-Risk TAVR Patients.. JACC. Cardiovascular interventions. 2020. DOI: 10.1016/j.jcin.2020.01.202.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.