BACKGROUND

Ampullary adenomas are increasingly being recognized, particularly in patients with familial adenomatous polyposis. A capsule endoscopy (CE) is routinely recommended for surveillance of small-intestinal polyposis. Performance characteristics of CE for the detection of periampullary lesions are unclear.

OBJECTIVE

To evaluate the ability of CE to detect the major duodenal papilla.

DESIGN AND PATIENTS: A total of 146 consecutive CE studies were reviewed by 2 CE gastroenterologists at 5 frames per second. Primary outcome was visualization of the major duodenal papilla. Discrepancies were reviewed by 5 CE gastroenterologists.

SETTING

A tertiary-referral center.

MAIN OUTCOME MEASUREMENTS: The ability of CE to detect the duodenal papilla.

RESULTS

Among 146 consecutive CE studies, 21 were excluded: capsule retention (3), patient age <18 years (6), duplicate study (8), and prior surgery disrupting duodenal anatomy (4). Of the remaining 125 studies, indications were the following: obscure GI bleeding (45.6%), iron deficiency anemia (19.2%), abdominal pain (17.6%), diarrhea (10.4%), and Crohn's disease (4.8%). In total, 13 major duodenal papillae were visualized. The median time of detection was 31 seconds after the first duodenal image. This translates to a CE sensitivity of 10.4% for detection of the major papilla.

LIMITATION: Papilla position was not verified by an EGD.

CONCLUSIONS

CE has limited sensitivity to visualize the major papilla and lesions in the periampullary small intestine. Nondiagnostic CE studies must not be relied upon as proof that small-bowel lesions do not exist. Consideration should be given for an enteroscopy or side-viewing duodenoscopy in cases where significant clinical concern exists for unrecognized periampullary lesions. The current recommendations about surveillance for small-bowel polyposis should be revised.

Для цитирования: Clarke John O, Giday Samuel A, Magno Priscilla, Shin Eun Ji, Buscaglia Jonathan M, Jagannath Sanjay B, Mullin Gerard E. How good is capsule endoscopy for detection of periampullary lesions? Results of a tertiary-referral center.. Gastrointestinal endoscopy. 2008. DOI: 10.1016/j.gie.2007.11.055.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.