Similarities and Differences Between the Extravascular and the Subcutaneous Implantable Cardioverter-Defibrillator: Clinical Selection and Implantation (Part 1).

Sugrue Alan, Friedman Paul

СтатьяОбзорПеревод в очереди

Аннотация

PURPOSE OF REVIEW: This review examines the two non-transvenous implantable cardioverter-defibrillator (ICD) platforms; the subcutaneous ICD (S-ICD) and the extravascular ICD (EV-ICD), providing a framework for patient selection, preimplant screening, and implantation.

RECENT FINDINGS: Long-term trial and registry data confirm that the S-ICD provides reliable defibrillation without intravascular leads, with lower rates of lead failure, infection, and pneumothorax compared to transvenous systems. The EV-ICD, utilizing a substernal lead, has demonstrated effective defibrillation and anti-tachycardia pacing in approximately 77% of episodes, while offering pause-prevention pacing in a smaller pulse generator with greater projected longevity. Device selection should be individualized based on pacing requirements, prior sternotomy status, and anticipated arrhythmia substrate. The S-ICD remains the only option after sternotomy, while the EV-ICD extends the non-transvenous paradigm by integrating anti-tachycardia and pause-prevention pacing. Together, these devices offer complementary solutions for patients who do not require chronic pacing or resynchronization therapy.

Другие публикации по теме

Defibrillators, ImplantablePatient SelectionDeath, Sudden, CardiacArrhythmias, Cardiac
Для цитирования: Sugrue Alan, Friedman Paul. Similarities and Differences Between the Extravascular and the Subcutaneous Implantable Cardioverter-Defibrillator: Clinical Selection and Implantation (Part 1).. Current cardiology reports. 2026. DOI: 10.1007/s11886-026-02403-2.