DOI 10.5603/rpor.a2022.0089

Медицинская библиотека · Reports of practical oncology and radiotherapy : journal of Greatpoland Cancer Center in Poznan and Polish Society of Radiation Oncology · 2022

Setup accuracy and dose attenuation of a wooden immobilization system for lung stereotactic body radiotherapy.

Nitta Yuya, Ueda Yoshihiro, Murata Seiya, Isono Masaru, Ohira Shingo, Masaoka Akira, Inui Shoki, Washio Hayate, Yoshinaka Reimi, Sagawa Tomohiro, Miyazaki Masayoshi, Teshima Teruki

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

We evaluated the setup error and dose absorption of an immobilization system with a shell and wooden baseplate (SW) for lung stereotactic body radiotherapy (SBRT).

MATERIALS AND

METHODS

Setup errors in 109 patients immobilized with an SW or BodyFix system (BF) were compared. Dose attenuation rates of materials for baseplates were measured with an ion-chamber. Ionization measurements were performed from 90° to 180° gantry angle in 10° increments, with the ball water equivalent phantom placed at the center of the wood and carbon baseplates whose effects on dose distribution were compared using an electron portal imaging device.

RESULTS

The ratio for the anterior-posterior, cranial-caudal, and right-left of the cases within 3-mm registered shifts in interfractional setup error were 90.9%, 89.2%, and 97.4% for the SW, and 93.2%, 91.6%, and 98.0% for the BF, respectively. For intrafractional setup error, 98.3%, 97.4%, and 99.1% for the SW and 96.6%, 95.8%, and 98.7% for the BF were within 3-mm registered shifts, respectively. In the center position, the average (minimum/maximum) dose attenuation rates from 90° to 180° for the wooden and carbon baseplates were 0.5 (0.1/2.8)% and 1.0 (-0.1/10.1)% with 6 MV, respectively. The gamma passing rates of 2%/2 mm for the wooden and carbon baseplates were 99.7% and 98.3% (p < 0.01).

CONCLUSIONS

The immobilization system with an SW is effective for lung SBRT since it is comparable to the BF in setup accuracy. Moreover, the wooden baseplate had lower radiation attenuation rates and affected the dose distribution less than the carbon baseplate.

Для цитирования: Nitta Yuya, Ueda Yoshihiro, Murata Seiya, Isono Masaru, Ohira Shingo, Masaoka Akira, Inui Shoki, Washio Hayate, Yoshinaka Reimi, Sagawa Tomohiro, Miyazaki Masayoshi, Teshima Teruki. Setup accuracy and dose attenuation of a wooden immobilization system for lung stereotactic body radiotherapy.. Reports of practical oncology and radiotherapy : journal of Greatpoland Cancer Center in Poznan and Polish Society of Radiation Oncology. 2022. DOI: 10.5603/rpor.a2022.0089.