Blood cell counting in neonates: a comparison between a low volume micromethod and the standard laboratory method.

Papa Fabrizio, Rongioletti Mauro, Ventura Marco Della, Di Turi Francesco, Cortesi Maurizio, Pasqualetti Patrizio, Majolini Maria Bernardetta, Collegiani Valeria, Cicchese Marika, Notarmuzi Maria Letizia, Agostino Rocco, Liumbruno Giancarlo Maria

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАВалидационное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

Iatrogenic anaemia caused by repeated blood sampling to monitor laboratory parameters can contribute, particularly in neonates, to the need for transfusion. "Point of care" laboratory equipment uses smaller amounts of blood for analytic determinations and could, therefore, help to prevent secondary anaemia. In this study we compared the results of haematological parameters measured using a standard laboratory method and using a "point of care" micromethod, with the aim of validating the use of this latter method in clinical practice in neonatology.

MATERIALS AND

METHODS

One hundred and fifty venous or capillary blood samples were taken from full-term or premature neonates 2-4 hours or 48 hours after birth. Each sample was processed by a standard haematology analyser and another micromethod instrument. Bland-Altman plots were constructed for each parameter and intra-class coefficients of correlation were calculated in order to evaluate the concordance between the two analysers.

RESULTS

The concordance between the data obtained with the two analysers, expressed as the intra-class correlation, was 0.98 for white blood cell count, 0.97 for haemoglobin concentration, 0.96 for haematocrit, 0.95 for mean red cell volume and 0.98 for platelet count. The micromethod produced overestimated mean values for the leucocyte count (+1.27; p<0.001), haematocrit (+1.80; p<0.001) and platelet count (+13.55; p<0.001).

CONCLUSIONS

Overall, the concordance between the values obtained with the two analysers was high for each of the parameters taken into consideration. In the case of haemoglobin and leucocytes, give the high intra-class correlation and lack of systematic overestimation of one method over another, the micromethod guarantees a correct evaluation; however, despite the high intra-class correlations for platelet counts, the systemic error seems to suggest that the micromethod cannot guarantee an appropriate evaluation of this parameter.

Для цитирования: Papa Fabrizio, Rongioletti Mauro, Ventura Marco Della, Di Turi Francesco, Cortesi Maurizio, Pasqualetti Patrizio, Majolini Maria Bernardetta, Collegiani Valeria, Cicchese Marika, Notarmuzi Maria Letizia, Agostino Rocco, Liumbruno Giancarlo Maria. Blood cell counting in neonates: a comparison between a low volume micromethod and the standard laboratory method.. Blood transfusion = Trasfusione del sangue. 2011. DOI: 10.2450/2011.0082-10.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.