Battery Lifespan of an Implantable Middle Ear Device.
Аннотация
BACKGROUND
The Carina system (Cochlear Ltd., Sydney, Australia) is a totally implantable device providing acoustic amplification in adult patients with moderate-to-severe sensorineural or mixed hearing loss. One of the main concerns about such a totally implantable device has been represented by the subcutaneous battery lifespan. The aim of this article is to report the analysis of battery performances in a series of Carina-implanted patients after a long follow up.
METHODS
In this retrospective study, the technical data of a series of patients implanted with the Carina middle ear implant in our clinic have been analysed, extracting the data from the log of telemetric measures.
RESULTS
The mean lifespan cutback was 0.43 h/years (from 0 to 0.71 h/year), with a strong negative significant correlation between the follow-up period and the percentage of battery residual lifespan.
CONCLUSION
The lifespan of the Carina's battery seems consistent with the manufacturer statement of a pluri-decennial lifespan, avoiding the need of an early surgical substitution and providing a full day of use of the system even after up to 12 years from the implantation.