MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1063/1.3030861

Медицинская библиотека · The Review of scientific instruments · 2008

Universal serial bus powered and controlled isolated constant-current physiological stimulator.

Holcomb Mark R, Bekele Robel Y, Lima Eduardo A, Wikswo John P

СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

We have developed a compact, isolated, physiological, constant-current stimulator that is powered and controlled by a universal serial bus (USB) interface. The stimulator is designed to be used in ex vivo cardiac experiments but is suitable for a wide variety of settings. The cost and features compare very favorably with commercial stimulators usually used in research and student laboratories. In addition to being USB powered, other novel aspects of our stimulator include the ability to produce large currents, up to 100 mA through a typical 1 kOmega load, by means of a single high-voltage dc-to-dc converter; user-specified variable period, magnitude, and duration of complex monophasic or biphasic sequences; and easy integration via hardware or software into existing experimental setups.

Для цитирования: Holcomb Mark R, Bekele Robel Y, Lima Eduardo A, Wikswo John P. Universal serial bus powered and controlled isolated constant-current physiological stimulator.. The Review of scientific instruments. 2008. DOI: 10.1063/1.3030861.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.