Accuracy and reliability comparison between different intraoral scanning devices in patients with permanent dentition: A prospective clinical study.

Cavalcanti Heloísa Nelson, Ribeiro de Almeida Lázaro Vinícius, da Silva Vinícius Merino, Bellini-Pereira Silvio Augusto, Janson Guilherme, Garib Daniela, Henriques José Fernando Castanha

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

This study primarily aimed to compare the accuracy and secondarily the reliability of different intraoral scanning (IOS) devices using intra- and interarch linear measurements of three-dimensional digital dental models.

METHODS

Digital dental models were divided into three groups based on their acquisition method: group T (GT) using the Trios® IOS, group P (GP) using the Panda® (IOS), and the control group (CG) consisting of plaster models digitized with the E3® desktop scanner. Two examiners measured the following variables using the OrthoAnalyzer® software: mesiodistal tooth width, clinical crown height, intercanine, interpremolar and intermolar distances, arch perimeter, arch length, overjet and overbite. Intra- and inter-examiner reliability was assessed with the intraclass correlation coefficient (ICC) and Bland-Altman test. Intergroup comparisons were performed using the analysis of variance (Anova) followed by Tukey tests (P<0.05).

RESULTS

The sample consisted of 25 volunteers (11 men, 14 women) with a mean age of 29.6years. Overall, the measurements of digital models obtained from the Trios®, Panda® and the E3® desktop scanner were comparable. For intermolar distance, the mean and standard deviation were as follows: GT (52.4±3.55), GP (52.6±3.56), and CG (52.1±3.96) with P=0.896 for the maxillary arch, and GT (45.8±3.87), GP (46.3±4.40), and CG (46.3±4.08), with P=0.848 for the mandibular arch. No statistically significant differences were observed regarding this and other variables. Measurements showed good to excellent reliability for both Trios® and Panda® IOS.

CONCLUSION

The three scanning devices Trios®, Panda® and E3® presented similar and adequate accuracy and reliability.

Для цитирования: Cavalcanti Heloísa Nelson, Ribeiro de Almeida Lázaro Vinícius, da Silva Vinícius Merino, Bellini-Pereira Silvio Augusto, Janson Guilherme, Garib Daniela, Henriques José Fernando Castanha. Accuracy and reliability comparison between different intraoral scanning devices in patients with permanent dentition: A prospective clinical study.. International orthodontics. 2025. DOI: 10.1016/j.ortho.2025.100983.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.