DOI 10.1093/dmfr/twaf005

Медицинская библиотека · Dento maxillo facial radiology · 2025

Assessment of the quality of root canal fillings-an ex vivo comparison of CBCT scans, conventional intraoral sensors, and a novel photon-counting sensor.

Jervis Matt, Waid Erin, Melo da Fonte Juliana B, Pita de Melo Daniela, Replogle Karan J, Sousa Melo Saulo L

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

To compare a novel photon-counting sensor, 2 CBCT protocols and 2 CMOS sensors on the detection of gaps between a gutta-percha cone and root canal walls.

METHODS

Twenty-five mandibular incisors were prepared to 45/0.04 (size/taper) at working length. Teeth were placed in a partially dentate mandible and single gutta-percha cones of 7 sizes were placed at length, one at a time, for image acquisition with a photon-counting sensor, 2 CBCT protocols (90 µm3, 120 µm3) and 2 CMOS sensors. Three calibrated observers assessed images for gap presence. Sensitivity, specificity, accuracy, AUC, and agreement with gold standard were determined using ANOVA and Tukey test (P ≤ .05).

RESULTS

Photon-counting sensor showed superior sensitivity and accuracy (88.47%, 81.57%), significantly higher than the CBCT protocols (50.70%-56.33%, 45.87%-53.17%). Contrarily, the photon-counting sensor showed the lowest specificity (40.27%), significantly lower than the CBCT protocols (90.27%, 97.23%). CMOS sensors showed sensitivity, specificity, and accuracy between 72.23%-74.53%, not differing from other modalities. All intraoral sensors showed AUC around 82.87%-84.03%, significantly higher than CBCT protocol 120 µm3 (74.07%). The file size was inversely related to gap size and percentual agreement with gold standard.

CONCLUSIONS

CMOS sensors showed consistent results, while the photon-counting sensor had the highest sensitivity but lacked specificity. CBCT protocols excelled in specificity but had lower sensitivity.

ADVANCES IN KNOWLEDGE: Novel photon-counting sensors and CBCT imaging provided no significant advantage over conventional sensors in assessing gaps as an indicator of quality of root canal filling. Furthermore, smaller gaps were more difficult to detect, regardless of the imaging technique used.

Для цитирования: Jervis Matt, Waid Erin, Melo da Fonte Juliana B, Pita de Melo Daniela, Replogle Karan J, Sousa Melo Saulo L. Assessment of the quality of root canal fillings-an ex vivo comparison of CBCT scans, conventional intraoral sensors, and a novel photon-counting sensor.. Dento maxillo facial radiology. 2025. DOI: 10.1093/dmfr/twaf005.