DOI 10.1080/14992027.2025.2568653

Медицинская библиотека · International journal of audiology · 2025

Towards an optimized monothermal caloric screening test: comparing stimuli and thresholds.

Özel Halil Erdem, Karakuzu Ahmet Taha, Temir Hümeyra, Alpay Muhammed, Çalışkan Sebla, Özdoğan Fatih, Genç Selahattin

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

OBJECTIVES

Monothermal caloric testing (MCT) shows potential as a screening method in vestibular assessment; however, the optimal stimulus modality and diagnostic threshold remain unclear. This study aimed to identify, within a single investigation, the most effective stimulus type (air or water; warm or cool) and the optimal cut-off threshold (15% or 25%) for maximising the diagnostic performance of MCT.

DESIGN

Retrospective study.

STUDY SAMPLE: Bithermal caloric test (BCT) results from 202 adults (103 water, 99 air) were analysed. MCT results were assessed at 15% and 25% thresholds based on sensitivity, specificity, and overall diagnostic accuracy, using BCT as the reference standard.

RESULTS

Warm stimuli demonstrated higher sensitivity, with warm air yielding the highest value (86.7%), while cool stimuli showed greater specificity, with cool water reaching the highest specificity (78.7%) at the 25% threshold. Warm air MCT resulted in the lowest false negative rate (6%) and highest diagnostic accuracy (94%), reducing the need for BCT to 57.6% of patients. Lowering the threshold to 15% slightly improved accuracy (95.9%) but increased BCT referrals (64.6%).

CONCLUSIONS

Warm air MCT appears to be an efficient screening tool for detecting unilateral vestibular weakness, offering high diagnostic accuracy while potentially reducing the need for comprehensive BCT.

Для цитирования: Özel Halil Erdem, Karakuzu Ahmet Taha, Temir Hümeyra, Alpay Muhammed, Çalışkan Sebla, Özdoğan Fatih, Genç Selahattin. Towards an optimized monothermal caloric screening test: comparing stimuli and thresholds.. International journal of audiology. 2025. DOI: 10.1080/14992027.2025.2568653.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.