DOI 10.1111/jce.14367

Медицинская библиотека · Journal of cardiovascular electrophysiology · 2020

The use of intracardiac echocardiography catheters in endocardial ablation of cardiac arrhythmia: Meta-analysis of efficiency, effectiveness, and safety outcomes.

Goya Masahiko, Frame Diana, Gache Larry, Ichishima Yoko, Tayar Daiane Oliveira, Goldstein Laura, Lee Stephanie Hsiao Yu

СтатьяМетаанализПоддержка исследования: вне правительства СШАСистематический обзорПеревод пока не подготовлен

Аннотация

AIMS: The optimal use of intracardiac echocardiography (ICE) may reduce fluoroscopy time and procedural complications during endocardial ablation of cardiac arrhythmias. Due to limited evidence in this area, we conducted the first systematic literature review and meta-analysis to evaluate outcomes associated with the use of ICE.

METHODS AND

RESULTS

Studies reporting the use of ICE during ablation procedures vs without ICE were searched using PubMed/MEDLINE. A meta-analysis was performed on the 19 studies (2186 patients) meeting inclusion criteria, collectively representing a broad range of arrhythmia mechanisms. Use of ICE was associated with significant reductions in fluoroscopy time (Hedges' g -1.06; 95% confidence interval [CI] -1.81 to -0.32; P < .01), fluoroscopy dose (Hedges' g -1.27; 95% CI -1.91 to -0.62; P < .01), and procedure time (Hedges' g -0.35; 95% CI -0.64 to -0.05; P = .02) vs ablation without ICE. A 6.95 minute reduction in fluoroscopy time and a 15.2 minute reduction in procedure time was observed between the ICE vs non-ICE groups. These efficiency gains were not associated with any decreased effectiveness or safety. Sensitivity analyses limiting studies to an atrial fibrillation (AF) only population yielded similar results to the main analysis.

CONCLUSION

The use of ICE in the ablation of cardiac arrhythmias is associated with significantly lower fluoroscopy time, fluoroscopy dose, and shorter procedure time vs ablation without ICE. These efficiency improvements did not compromise the clinical effectiveness or safety of the procedure.

Для цитирования: Goya Masahiko, Frame Diana, Gache Larry, Ichishima Yoko, Tayar Daiane Oliveira, Goldstein Laura, Lee Stephanie Hsiao Yu. The use of intracardiac echocardiography catheters in endocardial ablation of cardiac arrhythmia: Meta-analysis of efficiency, effectiveness, and safety outcomes.. Journal of cardiovascular electrophysiology. 2020. DOI: 10.1111/jce.14367.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.