Effective clinical applications of Monte Carlo-based independent secondary dose verification software for helical tomotherapy.

Kosaka Takahiro, Takatsu Jun, Inoue Tatsuya, Hara Naoya, Mitsuhashi Taira, Suzuki Michimasa, Shikama Naoto

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To investigate the scope of the effective clinical application of Monte Carlo (MC)-based independent dose verification software for helical tomotherapy.

METHODS

DoseCHECK was selected as the MC-based dose calculation software. First, the dose calculation accuracy of DoseCHECK was evaluated with film and chamber measurements in a water-equivalent phantom. Second, the dose calculation accuracy was examined in several heterogeneous materials. Finally, dosimetric comparisons between DoseCHECK and the treatment planning system (TPS) were performed for clinical patient plans. Prostate IMRT, head and neck IMRT (HN), total body irradiation (TBI), and brain stereotactic radiotherapy (SRT) were evaluated.

RESULT: The DoseCHECK calculations agreed with the chamber and film measurements in the homogenous phantom. For heterogeneous phantom cases, the dose differences between DoseCHECK and TPS were within 3 %, except in air, in which large dose differences of 20 % were observed. In clinical patient plans, the median dose differences between the lung Dmean in TBI cases and the normal brain Dmean in brain SRT cases were significantly >3 %. For HN and brain SRT cases, the median target dose differences were >3 %.

CONCLUSION

Our results show that independent dose verification with the MC algorithm can detect systematic errors caused by the lack of heterogeneity correction in the TPS. In particular, MC-based independent dose verification is required for HN, TBI, and brain SRT cases in helical tomotherapy.

Для цитирования: Kosaka Takahiro, Takatsu Jun, Inoue Tatsuya, Hara Naoya, Mitsuhashi Taira, Suzuki Michimasa, Shikama Naoto. Effective clinical applications of Monte Carlo-based independent secondary dose verification software for helical tomotherapy.. Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB). 2022. DOI: 10.1016/j.ejmp.2022.11.003.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.