MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.phro.2022.10.001

Медицинская библиотека · Physics and imaging in radiation oncology · 2022

Statistical evaluation of the effectiveness of dual amplitude-gated stereotactic body radiotherapy using fiducial markers and lung volume.

Tanabe Yoshinori, Tanaka Hidekazu

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND AND PURPOSE: The low tracking accuracy of lung stereotactic body radiotherapy (SBRT) risks reduced treatment efficacy. We used four-dimensional computed tomography (4DCT) images to determine the correlation between changes in fiducial marker positions and lung volume for lung tumors, and we evaluated the effectiveness of the combined use of these images in lung SBRT.

MATERIALS AND

METHODS

Data of 30 patients who underwent fiducial marker placement were retrospectively analyzed. We calculated the motion amplitudes of the center of gravity coordinates of the lung tumor and fiducial markers in each phase and the ipsilateral, contralateral, and bilateral lung volumes using 4DCT. Moreover, we calculated the cross-correlation coefficient between the fiducial marker position and the lung volume changes waveform for the motion amplitude waveform of the lung tumor over three gating windows (all phases, ≤2 mm3, and ≤3 mm3).

RESULTS

Compared with the lung volume, approximately 30 % of the fiducial markers demonstrated a low correlation with the lung tumor. In the ≤2 mm3 and ≤3 mm3 gating windows, the cross-correlation coefficients between the lung tumor and the optimal marker (r > 0.9: 83 % and 86 %) were significantly different for all fiducial markers (r > 0.9: 39 %, 53 %) and the ipsilateral (r > 0.9: 35 % and 40 %), contralateral (r > 0.9: 44 % and 41 %), and bilateral (r > 0.9: 39 % and 45 %) lung volumes.

CONCLUSIONS

Some of the fiducial markers showed a low correlation with the lung tumor. This study indicated that the combined use of lung volume monitoring can improve tracking accuracy.

Для цитирования: Tanabe Yoshinori, Tanaka Hidekazu. Statistical evaluation of the effectiveness of dual amplitude-gated stereotactic body radiotherapy using fiducial markers and lung volume.. Physics and imaging in radiation oncology. 2022. DOI: 10.1016/j.phro.2022.10.001.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.