Shibata Mayuri, Okada Wataru, Sano Keisuke, Nakamura Kenji, Sakai Yusuke, Ueda Yoshihiro, Uemoto Kenji, Takei Yoshiki, Tanooka Masao
Аннотация
PURPOSE: We aimed to investigate the usefulness of iViz air ver.4 Convex (FUJIFILM, Tokyo) as a tool to determine the bladder capacity before prostate radiotherapy by comparing it with the existing BladderScan BVI 6100 (Verathon Inc., Bothell, Washington).
METHODS: We investigated the usefulness of iViz air as a tool to determine the bladder capacity before prostate radiotherapy by comparing it with the current BladderScan.
RESULTS: The absolute value of each error rate was approximately 30.9%±27.2% and 26.4%±18.9% for the BladderScan and iViz air, respectively, with no significant differences between the instruments (p=0.16). Evaluated by urine volume, the mean error rates for bladder volumes >50 ml were 26.9%±19.0% and 26.1%±18.5% for the Bladder Scan and iViz air, respectively, with no significant differences (p=0.56). However, the BladderScan and iViz air had significantly higher error rates of 89.5%±52.5% and 31.5%±25.1%, respectively, if the bladder volume was <50 ml (p=0.005).
CONCLUSION: The iViz air has limited measurement error to confirm images, especially in limited volumes, suggesting that it is a useful bladder capacity measurement device in performing prostate radiotherapy.
METHODS: We investigated the usefulness of iViz air as a tool to determine the bladder capacity before prostate radiotherapy by comparing it with the current BladderScan.
RESULTS: The absolute value of each error rate was approximately 30.9%±27.2% and 26.4%±18.9% for the BladderScan and iViz air, respectively, with no significant differences between the instruments (p=0.16). Evaluated by urine volume, the mean error rates for bladder volumes >50 ml were 26.9%±19.0% and 26.1%±18.5% for the Bladder Scan and iViz air, respectively, with no significant differences (p=0.56). However, the BladderScan and iViz air had significantly higher error rates of 89.5%±52.5% and 31.5%±25.1%, respectively, if the bladder volume was <50 ml (p=0.005).
CONCLUSION: The iViz air has limited measurement error to confirm images, especially in limited volumes, suggesting that it is a useful bladder capacity measurement device in performing prostate radiotherapy.