[Usefulness of an Ultrasound System with Automatic Bladder Urine Volume Measurement Using Artificial Intelligence Technology in Radiotherapy].
Аннотация
PURPOSE: We aimed to investigate the usefulness of iViz air ver.4 Convex (FUJIFILM, Tokyo) as a tool to determine the bladder capacity before prostate radiotherapy by comparing it with the existing BladderScan BVI 6100 (Verathon Inc., Bothell, Washington).
METHODS
We investigated the usefulness of iViz air as a tool to determine the bladder capacity before prostate radiotherapy by comparing it with the current BladderScan.
RESULTS
The absolute value of each error rate was approximately 30.9%±27.2% and 26.4%±18.9% for the BladderScan and iViz air, respectively, with no significant differences between the instruments (p=0.16). Evaluated by urine volume, the mean error rates for bladder volumes >50 ml were 26.9%±19.0% and 26.1%±18.5% for the Bladder Scan and iViz air, respectively, with no significant differences (p=0.56). However, the BladderScan and iViz air had significantly higher error rates of 89.5%±52.5% and 31.5%±25.1%, respectively, if the bladder volume was <50 ml (p=0.005).
CONCLUSION
The iViz air has limited measurement error to confirm images, especially in limited volumes, suggesting that it is a useful bladder capacity measurement device in performing prostate radiotherapy.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.