DOI 10.1016/j.jpeds.2010.02.003
Continuous glucose monitoring in newborn babies at risk of hypoglycemia.
Аннотация
OBJECTIVE
To determine the usefulness of continuous glucose monitoring in babies at risk of neonatal hypoglycemia.
STUDY
DESIGN
Babies >/=32 weeks old who were at risk of hypoglycemia and admitted to newborn intensive care received routine treatment, including intermittent blood glucose measurement using the glucose oxidase method, and blinded continuous interstitial glucose monitoring.
RESULTS
Continuous glucose monitoring was well tolerated in 102 infants. There was good agreement between blood and interstitial glucose concentrations (mean difference, 0.0 mmol/L; 95% CI, -1.1-1.1). Low glucose concentrations (<2.6 mmol/L) were detected in 32 babies (32%) with blood sampling and in 45 babies (44%) with continuous monitoring. There were 265 episodes of low interstitial glucose concentrations, 215 (81%) of which were not detected with blood glucose measurement. One hundred seven episodes in 34 babies lasted >30 minutes, 78 (73%) of which were not detected with blood glucose measurement.
CONCLUSION
Continuous interstitial glucose monitoring detects many more episodes of low glucose concentrations than blood glucose measurement. The physiological significance of these previously undetected episodes is unknown.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.