MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1080/14992027.2021.1964042

Медицинская библиотека · International journal of audiology · 2021

Equivalent threshold sound pressure levels for click and CE-Chirp® stimuli delivered by the SnapPROBE™.

Simonsen Lisbeth Birkelund, Geisler Ole, Laugesen Søren

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVE

To estimate the equivalent threshold sound pressure levels (ETSPLs) for nine stimuli used today and in the future for universal newborn hearing screening and diagnostics delivered by the Interacoustics SnapPROBE™.

DESIGN

The Click and eight variants of the CE-Chirp® were included to establish ETSPLs when delivered by the SnapPROBE™. The short duration stimuli were presented at a 20/s repetition rate for about 1-2 s following the international standards for establishing ETSPLs.

STUDY SAMPLE: 50 ears from a total of 27 normal hearing young adult subjects (female/male = 16/11; age = 21-25 years, mean age = 23.1 years).

RESULTS

Peak-to-peak equivalent signal levels and root mean square signal levels (at 20/s repetition rate) for the nine stimuli are stated.

CONCLUSIONS

The ETSPLs are in alignment overall with those found in previous studies, while some levels are about 2-6 dB higher for this study compared to the Interacoustics Titan IOW probe and the ER-3A insert earphone. The results are relevant for present and future universal newborn hearing screening and diagnostics carried out with the SnapPROBE™ to ensure that correct stimulation levels are applied.

Для цитирования: Simonsen Lisbeth Birkelund, Geisler Ole, Laugesen Søren. Equivalent threshold sound pressure levels for click and CE-Chirp® stimuli delivered by the SnapPROBE™.. International journal of audiology. 2021. DOI: 10.1080/14992027.2021.1964042.