DOI 10.1002/mrm.30605

Медицинская библиотека · Magnetic resonance in medicine · 2025

Quantitative abdominal sodium MRI combined with 32-channel proton pTx MRI at 7 Tesla in a large field-of-view.

Scheipers Anna K, Grimm Johannes A, Losch Jana, Orzada Stephan, Fiedler Thomas M, Nagel Armin M, Schmitter Sebastian, Ladd Mark E, Platt Tanja

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To combine large field-of-view abdominal   1 H $$ {}^1\mathrm{H} $$ and quantitative sodium (   23 Na $$ {}^{23}\mathrm{Na} $$ ) MRI in the same position at 7 T to enable the quantification of the tissue sodium concentration via   23 Na $$ {}^{23}\mathrm{Na} $$ MRI and the anatomical assignment in   1 H $$ {}^1\mathrm{H} $$ images without repositioning in several tissues and organs at once.

METHODS

A custom-built   23 Na $$ {}^{23}\mathrm{Na} $$ birdcage coil and reference vial setup together with a 32-channel   1 H $$ {}^1\mathrm{H} $$ pTx array were employed at 7 T to allow dual-nuclei MRI in the same position for a field-of-view (FOV) of ( 400 × 400 × 400 ) mm 3 $$ \left(400\times 400\times 400\right)\kern0.3em {\mathrm{mm}}^3 $$ . T 1 $$ {T}_1 $$ relaxation effects in the reference vials and phantom were corrected, B 1 + $$ {B}_1^{+} $$ maps were measured, and B 1 - $$ {B}_1^{-} $$ maps were simulated to correct the acquired   23 Na $$ {}^{23}\mathrm{Na} $$ data in post-processing. These corrections were evaluated in a phantom and then applied in vivo in three healthy volunteers.

RESULTS

In the phantom, it was demonstrated that proton and quantitative sodium MR images share the same large FOV. Phantom measurements showed an improved sodium concentration accuracy after the performed corrections. Large FOV   1 H $$ {}^1\mathrm{H} $$ and quantitative   23 Na $$ {}^{23}\mathrm{Na} $$ in vivo MRI was demonstrated in three healthy volunteers.

CONCLUSION

This work shows the feasibility of combined   1 H $$ {}^1\mathrm{H} $$ and quantitative   23 Na $$ {}^{23}\mathrm{Na} $$ imaging at 7 T in a large FOV both under free breathing in ≤ 47 min $$ \le 47\kern0.3em \min $$ , laying the ground work for an accurate evaluation of the tissue sodium concentration in several organs at once.

Для цитирования: Scheipers Anna K, Grimm Johannes A, Losch Jana, Orzada Stephan, Fiedler Thomas M, Nagel Armin M, Schmitter Sebastian, Ladd Mark E, Platt Tanja. Quantitative abdominal sodium MRI combined with 32-channel proton pTx MRI at 7 Tesla in a large field-of-view.. Magnetic resonance in medicine. 2025. DOI: 10.1002/mrm.30605.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.