Aneurysm dome and vessel pressure measurements with coiling, stent assisted coiling and flow diversion.

Becker Timothy A, Lewis Kailey L, Berns Holly F, Robertson Sophia E, Clark Wyatt E, Wells Jesse C, Alnajrani Mohammed K, Rapoport Christopher, Barhouse Patrick, Ramirez-Velandia Felipe, Filo Jean, Young Michael, Muram Sandeep, Granstein Justin H, Ogilvy Christopher S

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Variability in long-term endovascular treatment outcomes for intracranial aneurysms has prompted questions regarding the effects of these treatments on aneurysm hemodynamics. Endovascular techniques disrupt aneurysmal blood flow and shear, but their influence on intra-aneurysmal pressure remains unclear. A better understanding of aneurysm pressure effects may aid in predicting outcomes and guiding treatment decisions.

METHODS

Medium and large aneurysm models with intramural pressure taps on the dome and parent artery were designed and 3D-printed with vessel-like physical properties from UV-cured materials. The models were connected to a comprehensive flow system consisting of a pulsatile pump and a viscosity-matched blood analog. The system provided physiological pressure and flow control. Real-time pressures were recorded in the aneurysm dome and parent artery during initial placement of coils, stents, flow diverters, and temporary balloons under simulated surgical conditions. Coiling, stent-assisted coiling, and flow diverter placement were performed in both aneurysm sizes. Temporary balloon placement was performed in a large aneurysm model.

RESULTS

Coiling resulted in 24-30% packing density and diminished intra-aneurysmal flow. Flow diverter placement reduced intra-aneurysmal flow with near complete flow interruption after placement of three consecutive devices across the aneurysm neck. Compared to untreated controls, real-time pressure measurements during coiling and flow diversion showed minimal changes (< 5%) in intra-aneurysmal pressures. Temporary balloon occlusion blocked the parent artery, increasing the pressure proximal to the site of occlusion (by 9%), and reducing the pressure distally (by 14%). This maneuver also dampened intra-aneurysmal pressure to the average distal vessel pressure measurement. Positive control aneurysm models were 3D-printed with a sealed, "healed" neck. These controls verified a sealed neck eliminates intra-aneurysmal pressure.

CONCLUSION

Findings quantified minimal changes in intra-aneurysmal pressure during and immediately post-coiling and flow diversion. Intra-aneurysmal flow disruption alone has negligible impact on intra-aneurysmal pressures.

Для цитирования: Becker Timothy A, Lewis Kailey L, Berns Holly F, Robertson Sophia E, Clark Wyatt E, Wells Jesse C, Alnajrani Mohammed K, Rapoport Christopher, Barhouse Patrick, Ramirez-Velandia Felipe, Filo Jean, Young Michael, Muram Sandeep, Granstein Justin H, Ogilvy Christopher S. Aneurysm dome and vessel pressure measurements with coiling, stent assisted coiling and flow diversion.. Acta neurochirurgica. 2025. DOI: 10.1007/s00701-024-06392-5.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.