DOI 10.12455/j.issn.1671-7104.240688
[Production Research and Risk Factor Analysis of Transfusion and Infusion Warmer Based on Real-World Data].
Аннотация
OBJECTIVE
To investigate the transfusion and infusion warmer manufacturers, combine the use failures to analyze adverse events, and provide support for enterprise risk management and clinical safe use.
METHODS
The sentinels from 7 manufacturing enterprises and 11 medical institutions that participated in Shandong Province's key monitoring program during the "14th Five-Year Plan" period were targeted. This was done by understanding the equipment's principles, structures, and quality control. Additionally, real-world data from January 2019 to December 2023 were collected to count adverse events.
RESULTS
During production, there are risks in switching power supply stability and solder joint firmness. Fifteen kinds of faults occurred during use, and common faults such as inability to heat, unable to turn on the machine, and bubbles in the infusion tube accounted for more than 80%.
CONCLUSION
There are many risk points and failures for transfusion and infusion warmers, so enterprises should improve processes and quality control to address risks, and medical institutions should formulate specifications and maintenance plans to provide targeted theoretical basis for supervision.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.