MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jse.2019.02.028

Медицинская библиотека · Journal of shoulder and elbow surgery · 2019

Rotator cuff tendon tissue cut-through comparison between 2 high-tensile strength sutures.

Owens Brett D, Algeri Joseph, Liang Vivian, DeFroda Steven

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

High-tensile strength sutures are known to cut through tendon tissue when used for rotator cuff and other tendon repairs, resulting in mechanical failure. The purpose of this study was to test a new suture and compare it with an established suture in a controlled laboratory setting.

METHODS

Two sutures, Dynacord and FiberWire, both USP size No. 2, were passed through fresh infraspinatus tendons from 7 matched pairs of ovine shoulders (14 shoulders). Samples underwent cyclic testing for 1000 cycles, and the amount of cheese-wire tissue damage (tendon cut-through) was recorded. A clinical failure was defined as greater than 5 mm of tissue cut-through.

RESULTS

The mean amount of tendon cut-through was 3.72 ± 1.14 mm in the FiberWire specimens and 2.69 ± 1.02 mm in the Dynacord group. The difference was statistically significant (P = .012). In the matched-pair analysis, more tendon cut-through was noted with FiberWire in 13 specimens whereas a greater amount was found in only 1 Dynacord specimen. The FiberWire specimens showed 2 instances of tissue tendon cut-through exceeding 5 mm, defined as a clinical failure.

CONCLUSIONS

In this cadaveric ovine rotator cuff tendon model, we found less tendon cut-through from Dynacord suture compared with FiberWire. In addition, 2 of the FiberWire specimens showed complete tendon cut-through. Future studies focusing on patient-reported outcomes and healing rates with different types of suture materials are needed.

Для цитирования: Owens Brett D, Algeri Joseph, Liang Vivian, DeFroda Steven. Rotator cuff tendon tissue cut-through comparison between 2 high-tensile strength sutures.. Journal of shoulder and elbow surgery. 2019. DOI: 10.1016/j.jse.2019.02.028.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.