Broad-linewidth sources result in a skin-tone bias in noninvasive optical measurement of oxygen saturation.

Blaney Giles, Fantini Sergio

СтатьяПеревод пока не подготовлен

Аннотация

SIGNIFICANCE: Pulse oximeters overestimate arterial oxygen saturation in patients with darker skin tones, potentially increasing rates of occult hypoxemia. The emission of broad-linewidth light-emitting diodes (LEDs) has been proposed as a possible cause for this melanin-dependent bias.

AIM: This letter utilizes a physics-based model to investigate if the spectral bandwidth of light sources contributes to melanin-dependent measurement biases.

APPROACH: Pulse oximetry measurements were simulated across varying melanin concentrations and source bandwidths using a hybrid diffusion theory and Monte Carlo model. Epidermal melanin absorption was modeled with transmissive filters on the surfaces of an infinite slab representing bulk tissue.

RESULTS

Simulations demonstrate that pulse oximeter calibration curves depend on melanin concentration when broad-linewidth LEDs are utilized. Conversely, calibration curves remain independent of melanin when employing monochromatic laser diodes (LDs).

CONCLUSION

Spectral coloring from broad-linewidth LEDs contributes to the skin-tone bias observed in pulse oximetry. Although updating calibration methods for different skin tones is a potential approach, adopting monochromatic LDs could alternatively eliminate this spectral-coloring contribution to the bias.

Для цитирования: Blaney Giles, Fantini Sergio. Broad-linewidth sources result in a skin-tone bias in noninvasive optical measurement of oxygen saturation.. Journal of biomedical optics. 2026. DOI: 10.1117/1.jbo.31.7.070501.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.