The Journal of surgical research · 2010
Blood flow in the foreign-body capsules surrounding surgically implanted subcutaneous devices.
СтатьяПоддержка NIH: внешняя программаПоддержка правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Bartoli Carlo R, Godleski John J
Аннотация
BACKGROUND: Surgically implanted devices initiate inflammatory mechanisms and wound healing events and result in the formation of a thick fibrotic capsule that surrounds the device. To investigate the foreign-body response to devices of clinically relevant size, we used microspheres to determine regional blood flow patterns in the foreign-body capsule (FBC) and surrounding subcutaneous tissue after device implantation.
MATERIALS AND METHODS: In 10 canines, we implanted 40 subcutaneous devices (polysulfone n = 20, silicone-coated n = 10, titanium n = 10). Via thoracotomy, animals were instrumented with left atrial and aortic vascular access catheters for serial microsphere injections and reference blood sampling. Regional blood flow was repeatedly determined in the FBC, subcutaneous fascia surrounding the FBC, and subcutaneous fascia distal to the surgical site up to 19 wk after device implantation (n = 55 determinations).
RESULTS: Compared with normal blood flow in subcutaneous fascia distal to the surgical site, blood flow increased in FBCs surrounding each device material (polysulfone P = 0.0035, silicone-coated P < 0.0001, titanium P < 0.0001). Additionally, blood flow increased in the subcutaneous fascia within half a centimeter of fibrous capsules encasing polysulfone (P = 0.0081) but not silicone (P = 0.3706) or titanium (P = 0.8160) devices. The time-course of measured blood flow changes within FBCs were similar for polysulfone and silicone but not for titanium.
CONCLUSIONS: Surgically implanted subcutaneous devices of clinically relevant size elicit increases in blood flow in the FBC as well as surrounding fascia. Device material may influence regional blood flow patterns.
MATERIALS AND METHODS: In 10 canines, we implanted 40 subcutaneous devices (polysulfone n = 20, silicone-coated n = 10, titanium n = 10). Via thoracotomy, animals were instrumented with left atrial and aortic vascular access catheters for serial microsphere injections and reference blood sampling. Regional blood flow was repeatedly determined in the FBC, subcutaneous fascia surrounding the FBC, and subcutaneous fascia distal to the surgical site up to 19 wk after device implantation (n = 55 determinations).
RESULTS: Compared with normal blood flow in subcutaneous fascia distal to the surgical site, blood flow increased in FBCs surrounding each device material (polysulfone P = 0.0035, silicone-coated P < 0.0001, titanium P < 0.0001). Additionally, blood flow increased in the subcutaneous fascia within half a centimeter of fibrous capsules encasing polysulfone (P = 0.0081) but not silicone (P = 0.3706) or titanium (P = 0.8160) devices. The time-course of measured blood flow changes within FBCs were similar for polysulfone and silicone but not for titanium.
CONCLUSIONS: Surgically implanted subcutaneous devices of clinically relevant size elicit increases in blood flow in the FBC as well as surrounding fascia. Device material may influence regional blood flow patterns.