MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/s00405-023-07822-3

Медицинская библиотека · European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery · 2023

Audiological outcome after stapes surgery in relation to prosthesis type.

Handke Vanessa, Agha-Mir-Salim Parwis, James Paul, Müller Alexander

СтатьяПеревод в очереди

Аннотация

PURPOSE: Different techniques are used to fix crimp and CliP® Piston stapes prostheses to the long process of the incus (LPI). The CliP® Piston provides a stiff connection in contrast to the static bended loop of the crimp prosthesis, which imitates the physiological incudostapedial joint (ISJ) and thereby potentially leads to different hearing outcome.

METHODS

In a retrospective single-center study of German-speaking one hundred and ninety patients who underwent stapes surgery CliP® Piston or crimp prostheses between the years of 2016 and 2019 by the same surgeon and in the same setting. Pre- and postoperative bone- (BC) and air-conduction (AC) pure-tone thresholds, pre- and postoperative air-bone gap (ABG) for 0.5, 1, 1.5, 2, 3, 4 kHz and the surgery time were examined.

RESULTS

The postoperative bone conduction thresholds were significantly lower in the frequencies between 0.5 and 3 kHz and the mean ABG was < 10 dB in most cases independent of the prosthesis used. Crimp prosthesis showed a significantly better closure of the ABG at 0.5 kHz.

CONCLUSIONS

The audiological outcome after stapes surgery is dependent on the type of prosthesis used, as reflected by the frequency-specific air-bone gap. The better ABG closure with the crimp prosthesis might be the result of the connection to the LPI imitating the physiological ISJ. The crimp prosthesis may be the better choice if use of hearing aids is expected postoperatively.

Для цитирования: Handke Vanessa, Agha-Mir-Salim Parwis, James Paul, Müller Alexander. Audiological outcome after stapes surgery in relation to prosthesis type.. European archives of oto-rhino-laryngology : official journal of the European Federation of Oto-Rhino-Laryngological Societies (EUFOS) : affiliated with the German Society for Oto-Rhino-Laryngology - Head and Neck Surgery. 2023. DOI: 10.1007/s00405-023-07822-3.