Simulation of dB/dt-Over-Electric Field Cardiac Magnetostimulation Safety Ratios in 75 Body Models and 18 Gradient Systems.
Аннотация
PURPOSE: The IEC 60601-2-33 standard provides consensus-based safety provisions for MRI equipment. Protection of patients against cardiac stimulation (CS) is based on limiting the maximum E-field induced by MRI gradient coils. In practice, this is achieved by imposing a conservative dB/dt threshold on any gradient waveform. The dB/dt-over-E-field conversion ratio currently used in IEC 60601-2-33 was derived in a homogeneous ellipsoid exposed to a uniform B-field and is 10 (T/s)*(V/m)-1. This limit is becoming increasingly restrictive in high performance clinical systems. We therefore evaluate dB/dt-over-E-field ratios in realistic body models and coils using state-of-the-art electromagnetic simulations.
METHODS
We performed two independent simulation studies in a total of 75 realistic body models and 13 commercial gradient systems and derived dB/dt-over-E-field ratios in the heart. We thresholded the E-field maps to mitigate the impact of staircasing artifacts in boundary voxels between the myocardium and the lungs.
RESULTS
Thresholding the E-field maps at the 99th percentile E-field value (E99) eliminates staircasing artifacts in both simulation studies. Study #1 predicts a larger range of dB/dt-over-E99 ratios (13-53 (T/s)*(V/m)-1) than study #2 (12-35 (T/s)*(V/m)-1). Despite differences in EM solvers, body models, coils, mesh resolution, and post-processing, both studies find similar worst-case ratios of dB/dt-over-E99 of 12-13 (T/s)*(V/m)-1.
CONCLUSION
Our simulations of dB/dt-over-E-field ratios for cardiac safety in MRI cover a large range of realistic clinical scenarios. An increase of the allowable dB/dt beyond the current CS limit in IEC 60601-2-33 may be feasible.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.