Pasick Luke J, Tong Jane Y, Benito Daniel A, Sargi Zoukaa, Anis Mursalin M
Аннотация
PURPOSE: An estimated 34 % of reported operating room fires involve the airway. Despite the inherent risks in otolaryngologic surgery, education regarding prevention and management of airway fires is limited in graduate medical training. One contributing factor is a lack of reporting of such rare events in our literature.
METHODS: The U.S. Food and Drug Administration's Manufacturer and User Facility Device Experience database was queried for reports of adverse events related to fires occurring during surgical procedures of the airway from January 1, 2010, to March 31, 2020.
RESULTS: 3687 reports were identified and 49 unique reports of airway fire were included. Sustained fires were described in 16 (32.7 %) reports and 33 (67.3 %) described transient flares. 2 fires extended beyond the airway and 9 (18.4 %) were noted to have occurred at the start of the case. Fires were reported most commonly during tonsillectomy (n = 22 [44.9 %]), vocal fold excision (n = 5 [10.2 %]), and adenoidectomy (n = 4 [8.2 %]). 46 reports attributed flare initiation to a specific element of the fire triangle. 16 patient and 2 operator injuries were reported. Saline washing was utilized in 7 (14.3 %) cases overall. Patients were extubated immediately in 2 (12.5 %) of the 16 reports of sustained fires. 0 mortalities were reported.
CONCLUSION: Airway fires were reported in a variety of upper airway procedures performed regularly by otolaryngologists. The triggering factor that led to fire was identified as a spark or char in about half of the reported cases, and only 2 reports described immediate removal of the endotracheal tube.
METHODS: The U.S. Food and Drug Administration's Manufacturer and User Facility Device Experience database was queried for reports of adverse events related to fires occurring during surgical procedures of the airway from January 1, 2010, to March 31, 2020.
RESULTS: 3687 reports were identified and 49 unique reports of airway fire were included. Sustained fires were described in 16 (32.7 %) reports and 33 (67.3 %) described transient flares. 2 fires extended beyond the airway and 9 (18.4 %) were noted to have occurred at the start of the case. Fires were reported most commonly during tonsillectomy (n = 22 [44.9 %]), vocal fold excision (n = 5 [10.2 %]), and adenoidectomy (n = 4 [8.2 %]). 46 reports attributed flare initiation to a specific element of the fire triangle. 16 patient and 2 operator injuries were reported. Saline washing was utilized in 7 (14.3 %) cases overall. Patients were extubated immediately in 2 (12.5 %) of the 16 reports of sustained fires. 0 mortalities were reported.
CONCLUSION: Airway fires were reported in a variety of upper airway procedures performed regularly by otolaryngologists. The triggering factor that led to fire was identified as a spark or char in about half of the reported cases, and only 2 reports described immediate removal of the endotracheal tube.