Feasibility and limitations of head MRI in patients with cochlear implants.
Аннотация
OBJECTIVE
Cochlear implants (CIs) were noncompatible with magnetic resonance imaging (MRI) initially; however, recently, implants have become available that are compatible with MRI without the need for magnet removal or bandage fixation. The images produced by MRI scans are sometimes deteriorated by artifacts and are not clinically useful. In this study, we discussed the size differences of such artifacts with respect to the imaging modality and sequences with their clinical validity.
METHODS
We performed a head MRI, using a head bandage and without magnet removal in five patients who underwent cochlear implantation at our department and analyzed the MRI findings.
RESULTS
Without magnet removal, diffusion-weighted images and T2 star-weighted images had larger artifacts and less useful images. T1-weighted images, T2-weighted images (T2WIs), T2-weighted fluid-attenuated inversion recovery (T2-FLAIR) images, and heavy T2WIs could evaluate the unimplanted side and middle of the head but had limited applicability on the CI side.
CONCLUSION
The characteristic features of MRI scan images vary with the method used as well as with the sequence, suggesting that the choice of MRI is largely determined on the basis of clinical feasibility and the requirement. Accordingly, we need to judge well in advance of imaging whether the images would be clinically relevant.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.