DOI 10.1183/23120541.00220-2022
Evaluation of aerosol drug delivery with concurrent low- and high-flow nasal oxygen.
Аннотация
QUESTION ADDRESSED BY STUDY: Administration of aerosol to patients receiving high-flow nasal oxygen (HFNO) ranges from concurrent aerosol delivery by mouthpiece to aerosol via cannula alone. This study examines the conditions to provide optimal aerosol delivery with low- or high-flow nasal oxygen with concurrent mouthpiece or through nasal cannula alone, and the impact on fugitive aerosols.
MATERIALS AND
METHODS
A vibrating mesh nebuliser delivered salbutamol via mouthpiece, aerosol holding chamber and nasal cannula to an adult head model simulating relaxed breathing. The mean±sd inhaled dose (%) was assayed from a filter distal to the trachea. Optical particle sizers were used to measure fugitive aerosol concentrations during aerosol delivery.
RESULTS
Concurrent low-flow nasal oxygen (LFNO) and aerosol delivery with a mouthpiece and aerosol holding chamber increased the inhaled dose (%) available, 31.44±1.33% when supplemented with 2 L·min-1 of nasal oxygen. Concurrent HFNO above 30 L·min-1 resulted in a lower inhaled dose (%) compared to aerosol delivered through HFNO alone. The addition of concurrent LFNO or HFNO resulted in no increase in aerosol levels in the test room.
ANSWER TO QUESTION POSED: Concurrent LFNO with a mouthpiece and aerosol holding chamber is an effective and safe means of aerosol delivery.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.