DOI 10.1016/j.otohns.2004.05.014
The fluorescence of thyroid tissue.
Аннотация
OBJECTIVE
To investigate the fluorescent characteristics of different thyroid tissues.
STUDY
DESIGN
Fresh ex vivo thyroid tissue was surfaced scanned with a fluorescence spectrophotometer. Resultant spectra were evaluated via interpeak ratios.
RESULTS
Interpeak ratios of all histologies were found to be significantly different from each other (P < 0.001). The diagnostic sensitivity and specificity of fluorescent spectroscopy was 86% and 91% for thyroid nodule capsule, 78% and 59% for papillary thyroid carcinoma, and 82% and 63% for normal thyroid, respectively.
CONCLUSION
Fluorescent spectroscopy can been used to identify significant differences in the fluorescent characteristics of various thyroid histologies.
SIGNIFICANCE: This study is the first investigation of the fluorescence of thyroid tissue. The diagnostic ability of fluorescent spectrometry may allow it to be used as a localization aid for fine needle aspiration, using optical fiber probes. Further investigations may enhance the sensitivity and specificity of fluorescent spectrometry allowing it to replace or compliment fine needle aspiration.
EBM RATING: B-3.