PloS one · 2021

Diagnostic accuracy of Panbio rapid antigen tests on oropharyngeal swabs for detection of SARS-CoV-2.

Клиническое исследованиеСтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Ngo Nsoga Marie Thérèse, Kronig Ilona, Perez Rodriguez Francisco Javier, Sattonnet-Roche Pascale, Da Silva Diogo, Helbling Javan, Sacks Jilian A, de Vos Margaretha, Boehm Erik, Gayet-Ageron Angèle, Berger Alice, Jacquerioz-Bausch Frédérique, Chappuis François, Kaiser Laurent, Schibler Manuel, Renzoni Adriana, Eckerle Isabella

Аннотация

BACKGROUND: Antigen-detecting rapid diagnostic tests (Ag-RDTs) for the detection of SARS-CoV-2 offer new opportunities for testing in the context of the COVID-19 pandemic. Nasopharyngeal swabs (NPS) are the reference sample type, but oropharyngeal swabs (OPS) may be a more acceptable sample type in some patients.

METHODS: We conducted a prospective study in a single screening center to assess the diagnostic performance of the Panbio™ COVID-19 Ag Rapid Test (Abbott) on OPS compared with reverse-transcription quantitative PCR (RT-qPCR) using NPS during the second pandemic wave in Switzerland.

RESULTS: 402 outpatients were enrolled in a COVID-19 screening center, of whom 168 (41.8%) had a positive RT-qPCR test. The oropharyngeal Ag-RDT clinical sensitivity compared to nasopharyngeal RT-qPCR was 81% (95%CI: 74.2-86.6). Two false positives were noted out of the 234 RT-qPCR negative individuals, which resulted in a clinical specificity of 99.1% (95%CI: 96.9-99.9) for the Ag-RDT. For cycle threshold values ≤ 26.7 (≥ 1E6 SARS-CoV-2 genomes copies/mL, a presumed cut-off for infectious virus), 96.3% sensitivity (95%CI: 90.7-99.0%) was obtained with the Ag-RDT using OPS.

INTERPRETATION: Based on our findings, the diagnostic performance of the Panbio™ Covid-19 RDT with OPS samples, if taken by a trained person and high requirements regarding quality of the specimen, meet the criteria required by the WHO for Ag-RDTs (sensitivity ≥80% and specificity ≥97%) in a high incidence setting in symptomatic individuals.

Предметные рубрики MeSH