MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jviromet.2021.114201

Медицинская библиотека · Journal of virological methods · 2021

On-field evaluation of a ultra-rapid fluorescence immunoassay as a frontline test for SARS-CoV-2 diagnostic.

Orsi Andrea, Pennati Beatrice Marina, Bruzzone Bianca, Ricucci Valentina, Ferone Diego, Barbera Paolo, Arboscello Eleonora, Dentone Chiara, Icardi Giancarlo

Оценочное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Viral RNA amplification by real-time RT-PCR still represents the gold standard for the detection of SARS-CoV-2, but the development of rapid, reliable and easy-to-perform diagnostic methods is crucial for public health, because of the need of shortening the time of result-reporting with a cost-efficient approach.

OBJECTIVES

The aim of our research was to assess the performance of FREND™ COVID-19 Ag assay (NanoEntek, South Korea) as a ultra-rapid frontline test for SARS-CoV-2 identification, in comparison with RT-PCR and another COVID-19 antigen fluorescence immunoassay (FIA).

STUDY

DESIGN

The qualitative FIA FREND™ test, designed to detect within 3 min the Nucleocapsid protein of SARS-CoV-2, was evaluated using nasopharyngeal swabs in Universal Transport Medium (UTM™, Copan Diagnostics Inc, US) from suspected COVID-19 cases who accessed the Emergency Room of the Ospedale Policlinico San Martino, Genoa, Liguria, Northwest Italy. Diagnostic accuracy was determined in comparison with SARS-CoV-2 RT-PCR and STANDARD F™ COVID-19 Ag FIA test (SD BIOSENSOR Inc., Republic of Korea).

RESULTS

In November 2020, 110 nasopharyngeal samples were collected consecutively; 60 resulted RT-PCR positive. With respect to RT-PCR results, sensitivity and specificity of FREND™ COVID-19 Ag test were 93.3 % (95 % CI: 83.8-98.2) and 100 % (95 % CI: 92.9-100), respectively. FREND™and STANDARD F™ COVID-19 Ag FIA assays showed a concordance of 96.4 % (Cohen's k = 0.93, 95 % CI: 0.86-0.99).

CONCLUSIONS

FREND™ FIA test showed high sensitivity and specificity in nasopharyngeal swabs. The assay has the potential to become an important tool for an ultra-rapid identification of SARS-CoV-2 infection, particularly in situations with limited access to molecular diagnostics.

Для цитирования: Orsi Andrea, Pennati Beatrice Marina, Bruzzone Bianca, Ricucci Valentina, Ferone Diego, Barbera Paolo, Arboscello Eleonora, Dentone Chiara, Icardi Giancarlo. On-field evaluation of a ultra-rapid fluorescence immunoassay as a frontline test for SARS-CoV-2 diagnostic.. Journal of virological methods. 2021. DOI: 10.1016/j.jviromet.2021.114201.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.