Lumipulse G SARS-CoV-2 Ag assay evaluation using clinical samples from different testing groups.

Menchinelli Giulia, Bordi Licia, Liotti Flora Marzia, Palucci Ivana, Capobianchi Maria Rosaria, Sberna Giuseppe, Lalle Eleonora, Romano Lucio, De Angelis Giulia, Marchetti Simona, Sanguinetti Maurizio, Cattani Paola, Posteraro Brunella

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

Compared to RT-PCR, lower performance of antigen detection assays, including the Lumipulse G SARS-CoV-2 Ag assay, may depend on specific testing scenarios.

METHODS

We tested 594 nasopharyngeal swab samples from individuals with COVID-19 (RT-PCR cycle threshold [Ct] values ≤ 40) or non-COVID-19 (Ct values >40) diagnoses. RT-PCR positive samples were assigned to diagnostic, screening, or monitoring groups of testing.

RESULTS

With a limit of detection of 1.2 × 104 SARS-CoV-2 RNA copies/mL, Lumipulse showed positive percent agreement (PPA) of 79.9% (155/194) and negative percent agreement of 99.3% (397/400), whereas PPAs were 100% for samples with Ct values of <18 or 18-<25 and 92.5% for samples with Ct values of 25-<30. By three groups, Lumipulse showed PPA of 87.0% (60/69), 81.1% (43/53), or 72.2% (52/72), respectively, whereas PPA was 100% for samples with Ct values of <18 or 18-<25, and was 94.4, 80.0, or 100% for samples with Ct values of 25-<30, respectively. Additional testing of RT-PCR positive samples for SARS-CoV-2 subgenomic RNA showed that, by three groups, PPA was 63.8% (44/69), 62.3% (33/53), or 33.3% (24/72), respectively. PPAs dropped to 55.6, 20.0, or 41.7% for samples with Ct values of 25-<30, respectively. All 101 samples with a subgenomic RNA positive result had a Lumipulse assay's antigen positive result, whereas only 54 (58.1%) of remaining 93 samples had a Lumipulse assay's antigen positive result.

CONCLUSIONS

Lumipulse assay was highly sensitive in samples with low RT-PCR Ct values, implying repeated testing to reduce consequences of false-negative results.

Для цитирования: Menchinelli Giulia, Bordi Licia, Liotti Flora Marzia, Palucci Ivana, Capobianchi Maria Rosaria, Sberna Giuseppe, Lalle Eleonora, Romano Lucio, De Angelis Giulia, Marchetti Simona, Sanguinetti Maurizio, Cattani Paola, Posteraro Brunella. Lumipulse G SARS-CoV-2 Ag assay evaluation using clinical samples from different testing groups.. Clinical chemistry and laboratory medicine. 2021. DOI: 10.1515/cclm-2021-0182.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.