Prolonged Unfrozen Storage and Repeated Freeze-Thawing of SARS-CoV-2 Patient Samples Have Minor Effects on SARS-CoV-2 Detectability by RT-PCR.
СтатьяПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Dzung Andreas, Cheng Phil F, Stoffel Corinne, Tastanova Aizhan, Turko Patrick, Levesque Mitchell P, Bosshard Philipp P
Аннотация
Reliable transportation of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) patient samples from a swabbing station to a diagnostics facility is essential for accurate results. Therefore, cooling or freezing the samples is recommended in case of longer transportation times. In this study, SARS-CoV-2 detectability by RT-PCR was assessed after prolonged unfrozen storage or repetitive freeze-thawing of SARS-CoV-2 samples. SARS-CoV-2-positive patient swabs stored in viral transport medium were exposed to different temperatures (4°C, 25°C, and 35°C) and to repetitive freeze-thawing, to assess the effect of storage conditions on RT-PCR detection. SARS-CoV-2 RNA was still reliably detected by RT-PCR after 21 days of storage in viral transport medium, even when the samples had been stored at 35°C. The maximum observed change in cycle threshold value per day was 0.046 (±0.019) at 35°C, and the maximum observed change in cycle threshold value per freeze-thaw cycle per day was 0.197 (±0.06). Compared with storage at 4°C, viral RNA levels deviated little but significantly when stored at 25°C or 35°C, or after repeated freeze-thawing. The results of this study indicate that viral RNA levels are relatively stable at higher temperatures and repetitive freeze-thawing.