DOI 10.1164/rccm.200507-1148oc
Quantitative upper airway imaging with anatomic optical coherence tomography.
Аннотация
BACKGROUND
Measurements of upper airway size and shape are important in investigating the pathophysiology of obstructive sleep apnea (OSA) and in devising, applying, and determining the effectiveness of treatment modalities. We describe an endoscopic optical technique (anatomic optical coherence tomography, aOCT) that provides quantitative real-time imaging of the internal anatomy of the human upper airway.
METHODS
Validation studies were performed by comparing aOCT- and computed tomography (CT)-derived measurements of cross-sectional area (CSA) in (1) conduits in a wax phantom and (2) the velo-, oro-, and hypopharynx during wakefulness in five volunteers. aOCT scanning was performed during sleep in one subject with OSA.
RESULTS
aOCT generated images of pharyngeal shape and measurements of CSA and internal dimensions that were comparable to radiographic CT images. The mean difference between aOCT- and CT-derived measurements of CSA in (1) the wax phantom was 2.1 mm(2) with limits of agreement (2 SD) from -13.2 to 17.4 mm(2) and intraclass correlation coefficient of 0.99 (p < 0.001) and (2) the pharyngeal airway was 14.1 mm(2) with limits of agreement from -43.7 to 57.8 mm(2) and intraclass correlation coefficient of 0.89 (p < 0.001). aOCT generated quantitative images of changes in upper airway size and shape before, during, and after an apneic event in an individual with OSA.
CONCLUSIONS
aOCT generates quantitative, real-time measurements of upper airway size and shape with minimal invasiveness, allowing study over lengthy periods during both sleep and wakefulness. These features should make it useful for study of upper airway behavior to investigate OSA pathophysiology and aid clinical management.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.